X-ray Tube Stator Compatibility Tables Supplement for CPI X-ray Generators P/N: 746026-00 Revision: AF CPI Canada Inc. Proprietary and Confidential The information contained in this document are proprietary to Communications & Power Industries Canada Inc. (CPI), have been designed and developed at private expense, and are the exclusive property of CPI. The information contained herein is loaned in confidence and may not be duplicated in whole or in part, or be used for re-engineering, reverse engineering, or otherwise reproducing in any form or creating or attempting to create or permitting, allowing or assisting others to create or manufacture CPI’s product or products derived therefrom, without the express written permission of CPI. Copyright © By Communications & Power Industries Canada Inc. All Rights Reserved. Table of Contents CPI Canada Inc Contents: Introduction ................................................................................ 3 Low-Speed Starter Tube Select Table ............................................. 4 Notes and Warnings ...................................................................................... 4 CMP 200® Generators ................................................................................ 4 CMP 150 ™ Generators ............................................................................... 4 Mobile DR Generators ................................................................................. 5 Indico IQ® Generators ............................................................................... 5 Indico 100® Generators ............................................................................. 6 Low-Speed Starter Tube Select Table – CMP 200® / CMP 150™ Generators ......... 7 Low-Speed Starter Tube Select Table – Mobile DR Generators .......................... 19 Low-Speed Starter Tube – Indico IQ® / Indico 100® Generators ...................... 23 Notes for Tables 1A, 1B, and 1C .................................................................... 38 Dual-Speed Starter Tube Select Table ........................................... 39 Note for Table 2 .......................................................................................... 59 Dual-Speed Starter and Indico IQ Low-Speed Starter X-ray Tube Codes ....................................................................................... 65 Notes and Warnings for Table 3 .................................................................... 68 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 1 Table of Contents CPI Canada Inc (This page intentionally left blank) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 2 Introduction CPI Canada Inc Introduction This supplement contains the X-Ray Tube Stator Compatibility Tables for the Low Speed Starter (LSS) and the Dual Speed Starter (DSS). It is recommended to refer to the following Supplement Guide for the proper X-ray tube service. Table 1A: Low-Speed Starter X-ray Tube Type Selection for CMP 150 / CMP 200 Generator and Notes and Warning Table 1B: Low-Speed Starter X-ray Tube Type Selection for Mobile DR Generator and Notes and Warning Table 1C: Low-Speed Starter X-ray Tube Type Selection for Indico IQ / Indico 100 Generator and Notes and Warning Table 2: Dual-Speed Starter X-ray Tube Type Selection for Indico IQ / Indico 100 / CMP 200 Generator and Notes for Table 2 Table 3: Dual-Speed Starter and Indico IQ LowSpeed Starter X-ray Tube Codes and Notes and Warnings Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 3 Low-Speed Starter X-ray Tube Select Table CPI Canada Inc Low-Speed Starter Tube Select Table This section contains notes and warnings for using low speed starter tube select table provided in this supplement. Notes and Warnings CMP 200® Generators The boost voltage may be set to either 240 VAC (factory default) or 120 VAC. This should be set to 240 VAC, except where specifically noted in Table 1A. The Low Speed Starter output is at the same frequency as the AC line (50 Hz or 60 Hz). The Low Speed Starter is integrated into the HV Auxiliary Board within the CMP 200® and CMP 200® DR generators. The phase-shift capacitor is chassis-mounted and may be changed in the field. The LSS part number corresponds to the value of capacitor installed. The boost time is software-configurable from the console or GenWare®. Note that these generators do not use a separate run voltage, instead cycling the boost voltage on/off as required. The boost duty cycle is not hardwareconfigurable. CMP 150 ™ Generators Warning: The boost voltage is the same as the AC line voltage and is not adjustable. Most X-ray tubes require a nominal 200 – 240 VAC supply. A few X-ray tubes are intended to be operated using a 100 – 120 VAC supply, as indicated within the tables. In some cases, operating at tube intended for 200-240 VAC at 100-120 VAC may be feasible, but requires increasing the boost time. If the alternate boost time is not indicated within the tables, contact the factory for assistance. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 4 Low-Speed Starter X-ray Tube Select Table CPI Canada Inc The boost time is software-configurable from the console or GenWare®. Note that these generators do not use a separate run voltage, instead cycling the boost voltage on/off as required. The boost duty cycle is not hardware-configurable. The Low Speed Starter output is at the same frequency as the AC line (50 Hz or 60 Hz). The Low Speed Starter is integrated into the Input-Multifunction-Room (IMR) Board within the CMP 150 generators. The phase-shift capacitor is chassismounted and may be changed in the field. The LSS part number corresponds to the value of capacitor installed. Mobile DR Generators The Low Speed Starter is a separate sub-assembly within Mobile DR generator configurations. The boost voltage is 240 VAC at 60 Hz, boost time is 1.5 seconds, and a 30 µF phase-shift capacitor is provided. Indico IQ® Generators The Low-Speed-Starter (LSS) synthesizes its output frequencies independently of the line frequency. It will operate most tubes at 60 Hz, as indicated in Table 1C. Unless indicated otherwise within Table 1C, setting the switches to the code indicated for the applicable Tube Type selects all required operating parameters. The Low Speed Starter part number corresponds to the configuration of phaseshift capacitors installed. Reference notes are provided at the end of Table 1C. Tube operating parameters for each switch code are contained in Table 3 for reference. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 5 Low-Speed Starter X-ray Tube Select Table CPI Canada Inc Indico 100® Generators The boost voltage may be set to either 240 VAC (factory default) or 120 VAC. This should be set to 240 VAC, except where specifically noted in Table 1C. The Low Speed Starter output is at the same frequency as the AC line (50 Hz or 60 Hz). The Low Speed Starter is a separate sub-assembly within the INDICO 100®Generator. The LSS part number corresponds to the value of phase-shift capacitor incorporated within this sub-assembly. The boost time is selectable to either 1.5 seconds or 2.5 seconds, and the run voltage is selectable to any of 52 VAC, 73 VAC or 94 VAC. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 6 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc Low-Speed Starter Tube Select Table – CMP 200® / CMP 150™ Generators Low-Speed-Starter X-ray Tube Types Table 1A: CMP 200® / CMP 150™ Generators Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 7 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME 240 VAC 1.5 sec 30 µF TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Chirana Rotax KA 125 (20/50 Ω stator) RIK-T 0.8/2.0 12/50 RIK-T 1.2/2.0 30/50 Comet DO7 / DX7 (25/50 Ω stator) DX7 DX71HS (See note 3) 240 VAC Comet DO9 / DX9 (20/50 Ω stator) DX9 DX91H / HS DX92H / HS DX93H / HS DX94HS DX96HS DX97HS (See note 3) DX10H / HS DX101H / HS DX104HS DX105HS DX106HS (See note 3) XSTAR8 (XST-8) XSTAR74 (XST-74) XSTAR-WHIS 240 VAC DA1002 (DU304-1.0/2.0) DA1029 (DU304-0.6/1.2) DA1036 (DU404-0.6/1.2) DA1083 (DU404-0.3/1.0) DA1092 (DU604-0.6/1.2) DA1094 (DU404-0.4/0.8) 240 VAC Comet D010 / DX10 (20/50 Ω stator) Comet XSTAR (25/50 Ω stator) Dunlee (Philips) DA10 series (20/50 Ω stator) 240 VAC 240 VAC 1.5 sec 1.5 sec 1.5 sec 1.5 sec 2.0 sec 30 µF 30 µF 30 µF 30 µF LSS PART NO. (See note 2) 903836-00 903836-00 903836-00 903836-00 903836-00 40 µF 903836-01 (50 or 60 Hz) (prefer) 30 µF 903836-00 (60 Hz only) CMP 150™ BOOST VOLTS ( AC SUPPLY) 120 VAC BOOST TIME N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC 240 VAC 120 VAC N/A 2.0 sec N/A 240 VAC 2.0 sec Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF SHIFT CAP. 30 µF LSS PART NO. (See note 2) 908306-00 30 µF 908306-00 30 µF 908306-00 30 µF 908306-00 30 µF 908306-00 40 µF 908306-01 (50 or 60 Hz) 30 µF (60 Hz only) 908306-00 Page 8 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) CMP 200® / CMP 200® DR TUBE TYPE (HOUSING) TUBE TYPE (INSERT) BOOST VOLTS BOOST TIME SHIFT CAP. LSS PART NO. (See note 2) Dunlee (Philips) DA90, DA90HS (36/44 Ω stator) (See note 4) DU 1750 DU 3050 DU 2550 DU 33100 (See note 6) DR1429 (DU304-0.6/1.2) DR1436 (DU404-0.6/1.2) DR1492 (DU604-0.6/1.2) DR1494 (DU404-0.4/0.8) 240 VAC 1.5 sec 30 µF 903836-00 Dunlee (Philips) DR1400 series (20/50 Ω stator) Dunlee (Philips) DR1800 series (19/22 Ω stator) (See note 4) DR1817 DR1833 Dunlee (Picker / Philips) PX1300 series (3” anode) “S” stator (15/30 Ω Dunlee (Picker / Philips) PX1400 series (4” anode) “S” stator (15/30 Ω) PX1302 (DU303-1.0/2.0) PX1312 (DU303-0.6/1.2) PX1351 (DU143-1.0/2.0) PX1402 PX1412 PX1415 PX1425 PX1429 PX1431 PX1436 PX1456 PX1463 PX1472 PX1473 PX1475 PX1482 PX1483 PX1492 PX1494 DU404 (0.6/1.25) 240 VAC 240 VAC 240 VAC 240 VAC 2.0 sec 1.5 sec 2.0 sec 2.0 sec CMP 150™ BOOST BOOST VOLTS TIME ( AC SUPPLY) 120 VAC N/A 240 VAC 1.5 sec 40 µF 903836-01 (50 or 60 Hz) (prefer) 120 VAC N/A 240 VAC 2.0 sec 30µF (60 Hz only) 903836-00 120 VAC N/A 240 VAC 2.0 sec 30 µF 903836-00 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 2.0 sec 120 VAC N/A 240 VAC 2.0 sec 30 µF 30 µF 903836-00 903836-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF SHIFT CAP. LSS PART NO. (See note 2) 30 µF 908306-00 40 µF 908306-01 (50 or 60 Hz) 30 µF 908306-00 (60 Hz only) 30 µF 908306-00 30 µF 908306-00 30 µF 908306-00 Page 9 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Dunlee (Picker / Philips) PX1400 series (4” anode) “S” stator (15/30 Ω) PX1402 (DU304-1.0/2.0) PX1412 (DU304-0.6/1.2) PX1415 (DU404-0.3/0.9) PX1425 PX1429 (DU304-0.6/1.2) PX1431 (DU304-0.7/1.4) PX1436 (DU404-0.6/1.2) PX1456 (DU404-0.3/1.0) PX1463 (DU604-0.6/1.2) PX1472 (DU604-0.3/1.0) PX1473 (DU404-0.6/1.2) PX1475 (DU754-0.6/1.2) PX1482 (DU404-0.6/1.0) PX1483 (DU404-0.3/1.0) PX1492 (DU604-0.6/1.2) PX1494 (DU404-0.4/0.8 240 VAC MX-75 240 VAC 2.0 sec 30 µF LSS PART NO. (See note 2) CMP 150™ BOOST VOLTS ( AC SUPPLY) BOOST TIME SHIFT CAP. LSS PART NO. (See note 2) 903836-00 120 VAC N/A 30 µF 908306-00 240 VAC 2.0 sec 120 VAC N/A 40 µF 908306-01 240 VAC 1.5 sec 120 VAC N/A 40 µF 908306-01 240 VAC 1.5 sec Eureka (see Varian) GE Maxiray 75 (3” anode) (23/23 Ω equal impedance “E” stator) * Dunlee tube GE Maxiray 100 (4” anode) (23/23 Ω equal impedance “E” stator) 1.5 sec 40 µF 903836-01 DU303 (1.0/2.0) * DU303 (0.6/1.2) * MX-100 240 VAC 1.5 sec 40 µF 903836-01 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 10 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Gilardoni Rotagil S/AS AR11-30 AR20-50 AR30-00-1 AR30-60 AR30-100 AR40-100 AR9000-1 AR9000-2 Hangzhou Kailong (see Kailong RADII) CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME 240 VAC 1.5 sec 30 µF Hangzhou Wandong WANRAY (Polygala / Yuan Zhi) XD51-20.40/125 (LQ16-XA2) XD51-10.40/125 (LQ16-XA4) XD52-30.50/125 (LQ17-XB1) XD57-20.50/150 (LQ16-XD57) IAE RTM 70H X20P C30, C31 X22 (25/62 Ω stator) X40 IAE X20P X22 C40 X40 (33/64 Ω stator) 240 VAC IAE C52, C52 Super C100 C352 (20/40 Ω stator) 240 VAC RTC 600HS RTC 700HS RTC 1000HS RTM 78H / HS RTM 90H / HS RTM 92H / HS RTM 101H / HS RTM 102H / HS RTM 780H / HS RTM 782H / HS X39 X40 X50H X50AH 240 VAC 240 VAC 1.5 sec 1.5 sec 1.5 sec 1.5 sec LSS PART NO. (See note 2) 903836-00 40 µF 903836-01 (50 Hz only) 30 µF 20 µF 903836-00 903836-02 CMP 150™ BOOST VOLTS ( AC SUPPLY) 120 VAC BOOST TIME N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec SHIFT CAP. 30 µF LSS PART NO. (See note 2) 908306-00 40 µF (50 Hz only) 908306-01 30 µF 908306-00 20 µF 908306-02 30 µF 903836-00 (prefer) 120 VAC N/A 30 µF 908306-00 (prefer) 40 µF 903836-01 240 VAC 1.5 sec 40 µF 908306-01 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 11 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME 30 µF 240 VAC 1.5 sec CMP 150™ BOOST VOLTS ( AC SUPPLY) 120 VAC BOOST TIME N/A 240 VAC 1.5 sec 30 µF 903836-00 (50 or 60 Hz) (prefer) 120 VAC N/A 240 VAC 1.5 sec 40 µF (50 Hz only) 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec N/A 1.8 sec N/A 1.8 sec N/A 30 µF 908306-00 (50 or 60 Hz) (prefer) 40 µF 908306-01 (50 Hz only) 903836-00 120 VAC 240 VAC 120 VAC 240 VAC 120 VAC 30 µF 908306-00 903836-00 120 VAC N/A 30 µF 908306-00 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Kailong RADII H1064 H1065 H1066 (25/62 Ω stator) KL64 KL65 KL66 Kailong RADII H1074 H1075 H1080 H1083 H1086 (20/50 Ω stator) KL74 KL75 KL80 KL80A KL83 KL86 Kailong RADII H1076 KL76 240 VAC 1.5 sec 20 µF Kailong RADII H2100 (20/50 Ω stator) KL100 240 VAC 1.8 sec 30 µF 903836-00 (50 or 60 Hz) (prefer) 903836-01 40 µF (50 Hz only) Machlett (Varex) Dynamax (DX) 52 “R” stator (16/50 Ω) Machlett (Varex) Dynamax (DX) 62 “R” stator (23/56 Ω) Dynamax (DX) 62U configured as “STD” or “R” stator (15/36 Ω) A-102 A-132 A-142 240 VAC 1.5 sec 30 µF A-192B A-196 240 VAC 1.5 sec 30 µF A-197 A-256 A-272 A-282 240 VAC 1.5 sec A-286 A-292 A-482 A-486 240 VAC 1.5 sec 30 µF LSS PART NO. (See note 2) 903836-00 903836-01 903836-02 30 µF Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF SHIFT CAP. 30 µF LSS PART NO. (See note 2) 908306-00 30 µF 908306-00 (50 or 60 Hz) (prefer) 40 µF (50 Hz only) 908306-01 20 µF 908306-02 Page 12 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Philips ROT350 ROT351 RO 1230 RO 1750 / DU 1750* RO 3050 / DU 3050* SRO 2050 SRO 2250 GD6 3050 ** PG-256 *** PG-292 *** RAD-34 *** RAD-50 *** (36/44 Ω stator) (See note 4) * Dunlee tube (See note 6) ** NAGO tube *** Varex (Varian) tube CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME 240 VAC 1.5 sec 30 µF LSS PART NO. (See note 2) 903836-00 CMP 150™ BOOST VOLTS ( AC SUPPLY) 120 VAC BOOST TIME N/A 240 VAC 1.5 sec 120 VAC (See note 10) 1.5 sec 240 VAC N/A 120 VAC (See note 10) 240 VAC 1.5 sec SHIFT CAP. 30 µF LSS PART NO. (See note 2) 908306-00 30 µF 908306-00 30 µF 908306-00 30 µF 908306-00 Picker (see Dunlee) Shimadzu CIRCLEX P18DE-85 RX80 RX81 RX82 (7.5/35 Ω stator) 120 VAC Shimadzu CIRCLEX P38E RX100 120 VAC Shimadzu CIRCLEX P18DK RX150 (8.4/24 Ω stator) 120 VAC 1.5 sec 1.5 sec 1.5 sec 30 µF 30 µF 30 µF 903836-00 903836-00 903836-00 120 VAC (See note 10) 240 VAC Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 N/A 1.5 sec N/A Rev. AF Page 13 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Siemens Opti-150 “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube Siemens (SXVT) RAY-6_1 RAY-8(S)_1 RAY-12(S)_1 (1-phase only) (20/50 Ω stator) 20/40 30/52R SG-796B * CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME 120 VAC 2.0 sec 40 µF SV 150/12/40-S SV 150/25/48-S SV 150/22/54-S SV 150/33/78R-S 240 VAC Siemens (SXVT) RAY-13(S)_1 RAY-14(S)_1 (1-phase only) (20/50 Ω stator) SV 150/33/78R-S 240 VAC Siemens (SXVT) SDR (1-phase only) (20/50 Ω stator) SDR 150/30/50-1 Siemens (SXVT) SV 150 “S” stator (1-phase only) (14/18 Ω) SV 150/40/80C-100_1 (See note 7) 240 VAC 120 VAC 1.5 sec 1.5 sec 1.5 sec 2.0 sec 30 µF 30 µF 30 µF 40 µF LSS PART NO. (See note 2) 903836-01 903836-00 903836-00 903836-00 903836-01 CMP 150™ BOOST VOLTS ( AC SUPPLY) 120 VAC (See note 10) BOOST TIME 2.0 sec 240 VAC N/A 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC (See note 10) 240 VAC 2.0 sec Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 SHIFT CAP. 40 µF LSS PART NO. (See note 2) 908306-01 30 µF 908306-00 30 µF 908306-00 30 µF 908306-00 40 µF 908306-01 N/A Rev. AF Page 14 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME CMP 150™ BOOST VOLTS ( AC SUPPLY) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba Rotanode Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing used. For ease of use, the Toshiba tubes are sorted by Tube Insert Type (“E” number) within this table. The installer must confirm the stator type for the planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the appropriate table entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same row within this table. 20 µF E7132 240 VAC 1.5 sec 20 µF 903836-02 120 VAC N/A 908306-02 E7239 E7240 E7242 240 VAC 1.5 sec (See note 8) Toshiba XH-121 XH-126 XH-150 XS-AV stator XS-RA stator (27.5/58 Ω) Toshiba XH-106V XH-180 XH-181 XS-AL stator (9.4/28.3 Ω) E7252 (See note 8) 120 VAC 1.5 sec 40 µF 903836-01 (50 or 60 Hz) (prefer) 30 µF (60 Hz only) Toshiba XH-106V XH-180 XH-181 XS-RA stator (27.5/58 Ω) E7252 (See note 8) Toshiba XH-157 XS-RB stator (20.2/38 Ω) E7254 E7255 (See note 8) 240 VAC 240 VAC 1.5 sec 1.5 sec LSS PART NO. (See note 2) 20 µF 30 µF 903836-00 903836-02 903836-00 BOOST TIME 120 VAC (See note 10) 240 VAC 1.5 sec 120 VAC (See note 10) 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 SHIFT CAP. LSS PART NO. (See note 2) 40 µF 903836-01 (prefer) 30 µF 908306-00 20 µF 908306-02 30 µF 908306-00 N/A N/A Rev. AF Page 15 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba XH-121 XH-126 XS-AV stator XS-RA stator (27.5/58 Ω) Toshiba XH-106V XH-180 XS-AL stator (9.4/28.3 Ω) E7299 (See note 8) Toshiba XH-121 XS-BA stator (18/47.5 Ω) E7843 (See note 8) Toshiba XH-121 XH-126 XH-150 XS-RA stator (27.5/58 Ω) Toshiba XH-112V-2 XS-AG stator (9.4/28.3 Ω) Toshiba XH-121 XS-AV stator (27.5/58 Ω) E7861 (See note 8) E7813 (See note 8) CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME 20 µF 240 VAC 1.5 sec 120 VAC 120 VAC 240 VAC 1.5 sec 2.5 sec 1.5 sec LSS PART NO. (See note 2) 903836-02 CMP 150™ BOOST VOLTS ( AC SUPPLY) 120 VAC BOOST TIME N/A 240 VAC 1.5 sec 1.5 sec SHIFT CAP. 20 µF LSS PART NO. (See note 2) 908306-02 40 µF (50 or 60 Hz) 903836-01 (prefer) 30 µF (60 Hz only) 903836-00 40 µF (50 or 60 Hz) 903836-01 (prefer) 120 VAC (See note 10) 240 VAC 120 VAC (See note 10) 240 VAC 120 VAC (See note 10) 30 µF (60 Hz only) 903836-00 240 VAC N/A 30 µF (60 Hz only) 908306-00 20 µF 903836-02 120 VAC N/A 20 µF 908306-02 240 VAC 1.5 sec N/A 1.5 sec N/A 1.5 sec 40 µF 903836-01 (50 or 60 Hz) (prefer) 30 µF (60 Hz only) 908306-00 903836-01 40 µF (50 or 60 Hz) (prefer) E7864 (See note 8) N/A N/A N/A N/A N/A N/A N/A N/A E7865 (See note 8) 240 VAC 1.5 sec 20 µF 903836-02 120 VAC N/A 20 µF 908306-02 240 VAC 1.5 sec Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 16 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba XH-112V-2 XS-AG stator (9.4/28.3 Ω) Toshiba XH-121 XS-AV-stator XS-RA stator (27.5/58 Ω) Toshiba XH-121 XH-126 XH-150 XS-AL stator (9.4/28.3 Ω) E7869 (See note 8) Toshiba XH-121 XH-126 XS-AV stator XS-RA stator (27.5/58 Ω) E7886 (See note 8) E7876 (See note 8) E7884 (See note 8) CMP 200® / CMP 200® DR BOOST BOOST SHIFT CAP. VOLTS TIME N/A N/A N/A 240 VAC 120 VAC 1.5 sec 1.5 sec 20 µF 1.5 sec CMP 150™ BOOST VOLTS ( AC SUPPLY) N/A BOOST TIME N/A SHIFT CAP. N/A LSS PART NO. (See note 2) N/A 903836-02 120 VAC N/A 20 µF 908306-02 240 VAC 1.5 sec 120 VAC (See note 10) 240 VAC 120 VAC (See note 10) 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 903836-01 40 µF (50 or 60 Hz) (prefer) 30 µF (60 Hz only) 240 VAC LSS PART NO. (See note 2) N/A 20 µF 903836-00 903836-02 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 N/A 1.5 sec 40 µF 908306-01 (50 or 60 Hz) (prefer) 30 µF (60 Hz only) 908306-00 20 µF 908306-02 N/A Rev. AF Page 17 Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators CPI Canada Inc TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Varex (Varian) B-100 “STD” stator (16/50 Ω) Varex (Varian) B-130 B-130H B-135H B-146 B-147 B-150 B-150H Std “R” stator (16/50 Ω) Varex (Varian / Eureka ) Diamond Emerald LEO Std “R” stator (20/50 Ω) Varex (Varian / Eureka) Sapphire Std “R” stator (20/50 Ω) CMP 200® / CMP 200® DR CMP 150™ BOOST VOLTS BOOST TIME SHIFT CAP. LSS PART NO. (See note 2) BOOST VOLTS ( AC SUPPLY) BOOST TIME SHIFT CAP. LSS PART NO. (See note 2) A-102 A-132 / A-134* A-142 / A-144* A-145 * (See note 9) A-152 A-182 / A-184* A-192 / A-194* A-195 A-196 A-197 A-272 / A-274* A-277 / A-278* A-282 / A-284* A-286 A-292 / A-294* A-482 G-242 G-256 G-292 / G-294* G-296 G-892 G-896 * (See note 9) RAD-8 RAD-12 RAD-13 RAD-14 RAD-68 240 VAC 1.5 sec 30 µF 903836-00 120 VAC N/A 30 µF 908306-00 240 VAC 1.5 sec 120 VAC N/A 30 µF 908306-00 240 VAC 1.5 sec 120 VAC N/A 240 VAC 1.5 sec 120 VAC N/A 30 µF 908306-00 240 VAC 1.5 sec RAD-21 RAD-40 RAD-44 RAD-56 RAD-60 RAD-92 RAD-94 240 VAC 120 VAC N/A 30 µF 908306-00 240 VAC 1.5 sec 240 VAC 240 VAC 240 VAC 1.5 sec 30 µF 903836-00 1.5 sec 1.5 sec 1.5 sec 30 µF 30 µF 903836-00 903836-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 18 Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators CPI Canada Inc Low-Speed Starter Tube Select Table – Mobile DR Generators Low-Speed-Starter X-ray Tube Types Table 1B: Mobile DR Generators Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 19 Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators CPI Canada Inc TABLE 1B: TUBE TYPES – LOW SPEED STARTER – MOBILE DR GENERATORS TUBE TYPE (HOUSING) TUBE TYPE (INSERT) BOOST VOLTS BOOST TIME SHIFT CAP. LSS PART NO. (See note 2) IAE C30, C31 (25/62 Ω stator) IAE C40 (33/64 Ω stator) X20P X22 X40 240 VAC 1.5 sec 30 µF 904832-00 X20P X22 X40 240 VAC 1.5 sec 30 µF 904832-00 Kailong RADII H1064, H1065, H1066 (25/62 Ω stator) KL64 KL65 KL66 240 VAC 1.5 sec 30 µF 904832-00 Kailong RADII H1074, H1075 H1080, H1083, H1086 (20/50 Ω stator) KL74, KL75 KL80, KL80A KL83, KL86 240 VAC 1.5 sec 30 µF 904832-00 Kailong RADII H1076 KL76 240 VAC 1.5 sec 30 µF 904832-00 Siemens (SXVT) RAY-6_1 RAY-8(S)_1 RAY-12(S)_1 (1-phase only) (20/50 Ω stator) SV 150/12/40-S SV 150/25/48-S SV 150/22/54-S SV 150/33/78R-S 240 VAC 1.5 sec 30 µF 904832-00 Siemens (SXVT) RAY-13(S)_1 RAY-14(S)_1 (1-phase only) (20/50 Ω stator) SV 150/12/40-S SV 150/25/48-S SV 150/22/54-S SV 150/33/78R-S 240 VAC 1.5 sec 30 µF 904832-00 Siemens (SXVT) SDR (1-phase only) (20/50 Ω stator) SDR 150/30/50-1 240 VAC 1.5 sec 30 µF 904832-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 20 Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators CPI Canada Inc TABLE 1B: TUBE TYPES – LOW SPEED STARTER – MOBILE DR GENERATORS (Continued) TUBE TYPE (HOUSING) Toshiba Rotanode Toshiba XH-106V, XH-180, XH-181 XS-RA stator (27.5/58 Ω) TUBE TYPE BOOST BOOST LSS PART NO. SHIFT CAP. (INSERT) VOLTS TIME (See note 2) Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing used. For ease of use, the Toshiba tubes are sorted by Tube Insert Type (“E” number) within this table. The installer must confirm the stator type for the planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the appropriate table entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same row within this table. 904832-00 30 µF E7252 240 VAC 1.5 sec (See note 1) Toshiba XH-121, XH-126, XH-150 XS-RA stator (27.5/58 Ω) E7861 (See note 1) 240 VAC 1.5 sec 30 µF 904832-00 Toshiba XH-121 XS-AV-stator XS-RA stator (27.5/58 Ω) Toshiba XH-121 XS-AV-stator XS-RA stator (27.5/58 Ω) Toshiba XH-121, XH-126 XS-AV stator XS-RA stator (27.5/58 Ω) E7865 (See note 1) 240 VAC 1.5 sec 30 µF 904832-00 E7876 (See note 1) 240 VAC 1.5 sec 30 µF 904832-00 E7886 (See note 1) 240 VAC 1.5 sec 30 µF 904832-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 21 Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators CPI Canada Inc TABLE 1B: TUBE TYPES – LOW SPEED STARTER – MOBILE DR GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) BOOST VOLTS BOOST TIME SHIFT CAP. LSS PART NO. (See note 2) Varex (Varian) Diamond Emerald LEO Std “R” stator (20/50 Ω) RAD-8 RAD-12 RAD-13 RAD-14 RAD-68 RAD-74 240 VAC 1.5 sec 30 µF 904832-00 NOTE 1: Complete Toshiba x-ray tube assemblies include the suffix “X, “FX”, “GX”, or “JX”, which are interchangeable with respect to rotor and anode characteristics. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 22 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc Low-Speed Starter Tube – Indico IQ® / Indico 100® Generators Low-Speed-Starter X-ray Tube Types Table 1C: Indico IQ® / Indico 100® Generators Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 23 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100®® GENERATORS INDICO IQ® CODE, OUTPUT SHIFT Switches FREQ. CAP. TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CGR (GE) Statorix 200A/240/260 (50/110 Ω stator) MN640 M641 MN641 M643 Type 28 (00111) 50 Hz CGR (GE) Statorix 550 (50/110 Ω stator) RN620 RN621 Type 28 (00111) Chirana Rotax KA 125 (20/50 Ω stator) RIK-T 0.8/2.0 12/50 RIK-T 1.2/2.0 30/50 Chirana Rotax KA 125 (20/20/20 Ω stator) INDICO 100® LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 12.5µF 905135-41 240 VAC 2.5 sec 94 VAC 12.5µF 732752-01 50 Hz 12.5µF 905135-41 240 VAC 2.5 sec 94 VAC 12.5µF 732752-01 Type 21 (10101) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 73 VAC 30 µF 732752-00 N/A N/A N/A N/A 240 VAC 1.5 sec 73 VAC 100µF, 52 mH 732752-04 (1........5) Comet DO7 / DX7 (25/50 Ω stator) DX7 DX71HS (See note 3) Type 0 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Comet DO9 / DX9 (20/50 Ω stator) DX9 DX91H / HS DX92H / HS DX93H / HS DX94HS DX96HS DX97HS (See note 3) Type 0 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 24 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Comet D010 / DX10 (20/50 Ω stator) DX10H / HS DX101H / HS DX104HS DX105HS DX106HS (See note 3) DX700HS (See note 3) Type 0 (00000) Comet XSTAR (25/50 Ω stator) XSTAR8 (XST-8) XSTAR74 (XST-74) XSTAR-WHIS Dunlee (Philips) DA10 series (20/50 Ω stator) Comet DO700 / DX700 (25/50 Ω stator) Dunlee (Philips) DA90, DA90HS (36/44 Ω stator) (See note 4) Dunlee (Philips) DA90, DA90HS (9/11 Ω stator) (See note 5) INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Type 1 (10000) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A Type 0 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 DA1002 (DU304-1.0/2.0) Type 4 DA1029 (DU304-0.6/1.2) (00100) DA1036 (DU404-0.6/1.2) DA1083 (DU404-0.3/1.0) DA1092 (DU604-0.6/1.2) DA1094 (DU404-0.4/0.8) DU404 (0.6/1.25) DU 1750 Type 9 DU 3050 (10010) DU 2550 DU 33100 (See note 6) DU 1750 Type 9 DU 3050 (10010) DU 2550 DU 33100 (See note 6) 60 Hz 33 µF 30 µF 40 µF 905135-41 905135-43 905135-44 240 VAC 2.5 sec 52 VAC 45 µF 732752-02 (50 or 60 Hz) (prefer) Switches (1........5) FREQ. 30 µF (60 Hz only) 732752-00 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 25 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Dunlee (Philips) DR1400 series (20/50 Ω stator) DR1429 (DU304-0.6/1.2) Type 4 DR1436 (DU404-0.6/1.2) (00100) DR1492 (DU604-0.6/1.2) DR1494 (DU404-0.4/0.8) Switches (1........5) INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. 60 Hz 33 µF 30 µF 40 µF 905135-41 905135-43 905135-44 240 VAC 2.5 sec 52 VAC 45 µF 732752-02 (50 or 60 Hz) (prefer) 30 µF 732752-00 (60 Hz only) FREQ. LSS PART NO. (See note 2) Dunlee (Philips) DR1817 DR1800 series DR1833 (19/22 Ω stator) (See note 4) Type 9 (10010) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Dunlee (Philips) DR1817 DR1800 series DR1833 (9/11 Ω stator) (See note 5) Type 9 (10010) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A PX1302 (DU303-1.0/2.0) Type 7 PX1312 (DU303-0.6/1.2) (11100) PX1351 (DU143-1.0/2.0) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 2.5 sec 52 VAC 30 µF 732752-00 PX1402 (DU304-1.0/2.0) Type 1 PX1412 (DU304-0.6/1.2) (10000) PX1415 (DU404-0.3/0.9) PX1429 (DU304-0.6/1.2) PX1431 (DU304-0.7/1.4) PX1436 (DU404-0.6/1.2) PX1456 (DU404-0.3/1.0) PX1463 (DU604-0.6/1.2) PX1472 (DU604-0.3/1.0) PX1473 (DU404-0.6/1.2) PX1475 (DU754-0.6/1.2) PX1482 (DU404-0.6/1.0) PX1483 (DU404-0.3/1.0) PX1492 (DU604-0.6/1.2) PX1494 (DU404-0.4/0.8) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 2.5 sec 52 VAC 30 µF 732752-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Dunlee (Picker / Philips) PX1300 series (3” anode) (“S” stator, 15/30 Ω) Dunlee (Picker / Philips) PX1400 series (4” anode) (“S” stator, 15/30 Ω) Page 26 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) INDICO IQ® CODE, OUTPUT Switches (1........5) FREQ. INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) PX1402 (DU304-1.0/2.0) Type 1 PX1412 (DU304-0.6/1.2) (10000) PX1415 (DU404-0.3/0.9) PX1425 PX1429 (DU304-0.6/1.2) PX1431 (DU304-0.7/1.4) PX1436 (DU404-0.6/1.2) PX1456 (DU404-0.3/1.0) PX1463 (DU604-0.6/1.2) PX1472 (DU604-0.3/1.0) PX1473 (DU404-0.6/1.2) PX1475 (DU754-0.6/1.2) PX1482 (DU404-0.6/1.0) PX1483 (DU404-0.3/1.0) PX1492 (DU604-0.6/1.2) PX1494 (DU404-0.4/0.8) Eureka (see Varex/Varian) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 2.5 sec 52 VAC 30 µF 732752-00 GE Maxiray 75 (3” anode) (23/23 Ω equal impedance “E” stator) Type 14 (01110) 60 Hz 45 µF 40 µF 905135-41 905135-44 240 VAC 1.5 sec 73 VAC 45 µF 732752-02 Type 20 (00101) 60 Hz 45 µF 40 µF 905135-41 905135-44 240 VAC 1.5 sec 73 VAC 45 µF 732752-02 Dunlee (Picker / Philips) PX1400 series (4” anode) “S” stator (15/30 Ω) * Dunlee tube GE Maxiray 100 (4” anode) (23/23 Ω equal impedance “E” stator) MX-75 DU303 (1.0/2.0) * DU303 (0.6/1.2) * MX-100 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 27 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) INDICO IQ® CODE, OUTPUT Switches (1........5) FREQ. INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) Gilardoni Rotagil S/AS AR11-30 AR20-50 AR30-00-1 AR30-60 AR30-100 AR40-100 AR9000-001 AR9000-002 Hangzhou Kailong (see Kailong RADII) Type 22 (01101) 50 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Hangzhou Wandong WANRAY (Polygala / Yuan Zhi) XD51-20.40/125 (LQ16-XA2) XD51-10.40/125 (LQ16-XA4) XD52-30.50/125 (LQ17-XB1) XD57-20.50/150 (LQ16-XD57) IAE RTM 70H X20P C30, C31 X22 (25/62 Ω stator) X40 IAE X39 X40 C40 X42 (33/64 Ω stator) Type 0 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 45 µF (50 Hz only) 732752-02 Type 10 (01010) 50 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Type 27 (11011) 50 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 (prefer) 45 µF 732752-02 IAE C52, C52 Super C100 C352 (20/40 Ω stator) RTC 600HS RTC 700HS RTC 1000HS RTM 78H / HS RTM 90H / HS RTM 92H / HS RTM 101H / HS RTM 102H / HS RTM 780H / HS RTM 782H / HS X39 X40 X50H X50AH Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 28 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT INDICO 100® TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Kailong RADII H1064 H1065 H1066 (25/62 Ω stator) KL64 KL65 KL66 Type 0 (00000) Kailong RADII H1074 H1075 H1080 H1083 H1086 (20/50 Ω stator) KL74 KL75 KL80 KL80A KL83 KL86 Type 0 (00000) Kailong RADII H1076 KL76 Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 240 VAC 1.5 sec Kailong RADII H2100 (20/50 Ω stator) KL100 Type 21 (10101) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 2.5 sec Machlett (Varex) Dynamax (DX) 52 “R” stator (16/50 Ω) Machlett (Varex) Dynamax (DX) 62 “R” stator (23/56 Ω) Dynamax (DX) 62U configured as “STD” or “R” stator (15/36 Ω) A-102 A-132 A-142 Type 0 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec A-192B A-196 A-197 A-256 A-272 A-282 A-286 A-292 A-482 A-486 Type 0 (00000) Type 1 (10000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec Switches (1........5) SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF (50 or 60 Hz) 732752-00 (prefer) 45 µF (50 Hz only) 732752-02 52 VAC 30 µF 732752-00 73 VAC 30 µF (50 or 60 Hz) 732752-00 (prefer) 732752-02 52 VAC 45 µF (50 Hz only) 30 µF 52 VAC 30 µF 732752-00 FREQ. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF 732752-00 Page 29 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Philips ROT350 ROT351 RO 1230 RO 1750 / DU 1750* RO 3050 / DU 3050* SRO 2050 SRO 2250 GD6 3050 ** PG-256 *** PG-292 *** RAD-34 *** RAD-50 *** RO 1230 RO 1750 / DU 1750* RO 3050 / DU 3050* SRO 2050 SRO 2250 GD6 3050 ** PG-256 *** PG-292 *** RAD-34 *** RAD-50 *** SRM 0310 (36/44 Ω stator) (See note 4) * Dunlee tube (See note 6) ** NAGO tube *** Varex (Varian) tube Philips ROT350 ROT351 (9/11 Ω stator) (See note 5) * Dunlee tube (See note 6) ** NAGO tube *** Varex (Varian) tube Philips ROT500 ROT501 (9/11 Ω stator) INDICO IQ® CODE, OUTPUT Switches (1........5) FREQ. INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) Type 9 (10010) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Type 9 (10010) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A Type 30 (01111) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A Picker (see Dunlee) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 30 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Shimadzu CIRCLEX RX80 RX81 RX82 (7.5/35 Ω stator) P18DE-85 Type 16 (00001) Shimadzu CIRCLEX RX100 P38E Shimadzu CIRCLEX RX150 (8.4/24 Ω stator) INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 60 Hz 33 µF 30 µF 905135-41 905135-43 120 VAC 1.5 sec (See note 10) 52 VAC 30 µF 732752-00 Type 16 (00001) 60 Hz 33 µF 30 µF 905135-41 905135-43 120 VAC 1.5 sec (See note 10) 52 VAC 30 µF 732752-00 P18DK Type 16 (00001) 60 Hz 33 µF 30 µF 905135-41 905135-43 120 VAC 1.5 sec (See note 10) 52 VAC 30 µF 732752-00 Siemens Opti-150 “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube Siemens SV 150 “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube Siemens (SXVT) RAY-6_1 RAY-8(S)_1 RAY-12(S)_1 (1-phase only) (20/50 Ω stator) 20/40 30/52R SG-796B * N/A N/A N/A N/A 120 VAC 2.5 sec (See note 10) 52 VAC 45 µF 732752-02 SV 150/40/80C-100L Type 23 SG-256B * (11101) 50 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A SV 150/12/40-S SV 150/25/48-S SV 150/22/54-S SV 150/33/78R-S Type 31 (11111) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 73 VAC 30 µF 732752-00 Siemens (SXVT) RAY-13(S)_1 RAY-14(S)_1 (1-phase only) (20/50 Ω stator) SV 150/33/78R-S Type 31 (11111) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 73 VAC 30 µF 732752-00 Switches (1........5) FREQ. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 BOOST TIME Rev. AF Page 31 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Siemens (SXVT) SDR (1-phase only) (20/50 Ω stator) SDR 150/30/50-1 Siemens (SXVT) SV 150 “S” stator (1-phase only) (14/18 Ω) Toshiba Rotanode SV 150/40/80C-100_1 Toshiba XH-112V-2 XH-112W XS-AA stator XS-AG stator (9.4/28.3 Ω) Toshiba XH-112V XS-AG stator (9.4/28.3 Ω) Toshiba XH-121 XH-126 XH-150 XS-AV stator XS-RA stator (27.5/58 Ω) (See note 7) INDICO IQ® CODE, OUTPUT Switches FREQ. (1........5) Type 31 60 Hz (11111) Type 23 (11101) 50 Hz INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 73 VAC 30 µF 732752-00 33 µF 30 µF 905135-41 905135-43 120 VAC 2.5 sec (See note 10) 52 VAC 45 µF 732752-02 Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing used. For ease of use, the Toshiba tubes are sorted by Tube Insert Type (“E” number) within this table. The installer must confirm the stator type for the planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the appropriate table entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same row within this table. N/A N/A 33 µF DRX-3624H Type 17 60 Hz 905135-41 N/A N/A N/A DRX-3724H 30 µF (10001) 905135-43 DRX-4634H E7100 Type 17 (10001) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A E7132 E7239 E7240 E7242 (See note 8) Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 32 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT INDICO 100® TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba XH-112V-2 XS-AA stator XS-AG stator (9.4/28.3 Ω) Toshiba XH-106V XH-180 XH-181 XS-AL stator (9.4/28.3 Ω) E7250 (See note 8) Type 17 (10001) E7252 (See note 8) Type 29 (10111) Toshiba XH-106V XH-180 XH-181 XS-RA stator (27.5/58 Ω) E7252 (See note 8) Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 240 VAC 1.5 sec Toshiba XH-157 XS-RB stator (20.2/38 Ω) E7254 E7255 (See note 8) Type 21 (10101) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC Toshiba XH-121 XH-126 XS-AV stator XS-RA stator (27.5/58 Ω) Toshiba XH-106V XH-180 XS-AL stator (9.4/28.3 Ω) E7299 (See note 8) Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 240 VAC E7813 (See note 8) Type 29 (10111) 60 Hz 45 µF 40 µF 905135-41 905135-44 Switches (1........5) SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A 60 Hz 45 µF 40 µF 905135-41 905135-44 120 VAC 1.5 sec (See note 10) 52 VAC 45 µF (50 or 60 Hz) 732752-02 (prefer) 30 µF (60 Hz only) 732752-00 94 VAC 30 µF 732752-00 1.5 sec 52 VAC 30 µF (60 Hz only) 732752-00 1.5 sec 52 VAC 30 µF 732752-00 1.5 sec 120 VAC (See note 10) 52 VAC 45 µF (50 or 60 Hz) 732752-02 (prefer) 30 µF (60 Hz only) 732752-00 FREQ. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 33 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT INDICO 100® TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba XH-121 XS-BA stator (18/47.5 Ω) E7843 (See note 8) Toshiba XH-121 XH-126 XH-150 XS-RA stator (27.5/58 Ω) Toshiba XH-112V-2 XS-AG stator (9.4/28.3 Ω) Toshiba XH-121 XS-AV stator (27.5/58 Ω) Toshiba XH-112V-2 XS-AG stator (9.4/28.3 Ω) Toshiba XH-121 XS-AV-stator XS-RA stator (27.5/58 Ω) E7861 (See note 8) Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 240 VAC 1.5 sec E7864 (See note 8) Type 17 (10001) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A E7865 (See note 8) Type 19 (11001) 60 Hz 25 µF 20 µF 905135-41 905135-42 E7869 (See note 8) Type 17 (10001) 60 Hz 33 µF 30 µF E7876 (See note 8) Type 19 (11001) 60 Hz 25 µF 20 µF Switches (1........5) Type 0 (00000) FREQ. 60 Hz SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 33 µF 30 µF 905135-41 905135-43 120 VAC 2.5 sec (See note 10) 52 VAC 45 µF (50 or 60 Hz) 732752-02 (prefer) 30 µF (60 Hz only) 732752-00 94 VAC 30 µF 732752-00 N/A N/A N/A N/A 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 905135-41 905135-43 N/A N/A N/A N/A N/A 905135-41 905135-42 240 VAC 1.5 sec 94 VAC 30 µF 732752-00 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 BOOST TIME Rev. AF Page 34 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba XH-121 XH-126 XH-150 XS-AL stator (9.4/28.3 Ω) E7884 (See note 8) Toshiba XH-121 XH-126 XS-AV stator XS-RA stator (27.5/58 Ω) E7886 (See note 8) INDICO IQ® CODE, OUTPUT Switches (1........5) Type 29 (10111) Type 19 (11001) FREQ. 60 Hz 60 Hz INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS 45 µF 40 µF 905135-41 905135-44 1.5 sec 120 VAC (See note 10) 25 µF 20 µF 905135-41 905135-42 240 VAC Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 BOOST TIME 1.5 sec Rev. AF RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 52 VAC 45 µF (50 or 60 Hz) 732752-02 (prefer) 30 µF (60 Hz only) 732752-00 30 µF 732752-00 94 VAC Page 35 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Varex (Varian) B-100 “STD” stator (16/50 Ω) A-102 A-132 / A-134* A-142 / A-144* A-145 * (See note 9) A-152 A-182 / A-184* A-192 / A-194* A-195 A-196 A-197 * (See note 9) A-272 / A-274* A-277 / A-278* A-282 / A-284* A-286 A-292 / A-294* A-482 G-242 G-256 G-292 / G-294* G-296 G-892 G-896 * (See note 9) G-1077 G-1078 G-1080 G-1082 / G-1084* G-1086 / G-1087* G-1092 / G-1094* * (See note 9) Varex (Varian) B-130 B-130H B-135H B-146 B-147 B-150 B-150H Std “R” stator (16/50 Ω) Varex (Varian) B-160 B-160H B-165H “R” Stator (16/50 Ω) INDICO IQ® CODE, OUTPUT Switches (1........5) FREQ. INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) Type 2 (01000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Type 2 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A Type 1 (10000) Type 15 (11110) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 36 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued) INDICO IQ® CODE, OUTPUT TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Varex (Varian) B-180 B-180H B-185 B-185H “R” Stator (16/50 Ω) Varex (Varian) B-199 “S” Stator (15/18 Ω) Varex (Varian/ Eureka) Diamond Emerald LEO Std “R” stator (20/50 Ω) Varex (Varian / Eureka) Sapphire Std “R” stator (20/50 Ω) G-1582 G-1592 G-1593 Type 15 (11110) SG-1096 INDICO 100® SHIFT CAP. LSS PART NO. (See notes 1,2) BOOST VOLTS BOOST TIME RUN VOLTS SHIFT CAP. LSS PART NO. (See note 2) 60 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A Type 18 (01001) 50 Hz 33 µF 30 µF 905135-41 905135-43 N/A N/A N/A N/A N/A RAD-8 RAD-12 RAD-13 RAD-14 RAD-68 RAD-74 Type 0 (00000) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 RAD-21 RAD-40 RAD-44 RAD-56 RAD-60 RAD-92 RAD-94 Type 5 (10100) 60 Hz 33 µF 30 µF 905135-41 905135-43 240 VAC 1.5 sec 52 VAC 30 µF 732752-00 Switches (1........5) FREQ. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 37 Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators CPI Canada Inc Notes for Tables 1A, 1B, and 1C NOTE 2: Where more than one Low Speed Starter part number is referenced for a particular tube type and generator combination, any of these is acceptable. In some cases, a preferred part number is identified. NOTE 3: Comet tube inserts with the prefix “DI” and “DX” are interchangeable. NOTE 4: Philips / Dunlee housing with the windings connected in series (low-speed / high impedance configuration). NOTE 5: Philips / Dunlee housing with the windings connected in parallel (highspeed / low impedance configuration). NOTE 6: Dunlee tube inserts with the prefix “DU” and Philips tube inserts with the prefix “RO” are interchangeable. Select the corresponding Philips insert type within the Generator software if the Dunlee type is not available. NOTE 7: Siemens (SXVT) housing with the “phase-shifted” windings connected in parallel (low impedance, high speed configuration). NOTE 8: Complete Toshiba x-ray tube assemblies include the suffix “X, “FX”, “GX”, or “JX”, which are interchangeable with respect to rotor and anode characteristics. NOTE 9: These X-ray tubes incorporate a control grid. Grid control is currently not supported by standard CPI generators. Connect the grid connection to Cathode Common when using these tubes, and select the insert type within the Generator software corresponding to the equivalent non-grid tube. NOTE 10: The starter boost voltage required is 120 VAC. This is not available with CMP 150™ High Frequency X-ray generators operating from a nominal 200-240V AC line. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 38 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc Dual-Speed Starter Tube Select Table Dual-Speed-Starter X-ray Tube Types Table 2: Indico IQ® / Indico 100® / CMP 200® Generators Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 39 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc The Dual Speed Starter (DSS) synthesizes its output frequencies independently of the line frequency will operate all tubes at 60/180 Hz, unless a particular tube only has published ratings to operate at 150 Hz. In this case, the starter will output 50/150 Hz when set for these tubes. Unless indicated otherwise within Table 2, setting the switches to the code indicated for the applicable Tube Type selects all required operating parameters. The Dual Speed Starter part number corresponds to the AC line voltage (except Indico IQ) and the configuration of phase-shift capacitors installed. Reference notes are provided at the end of Table 2. Tube operating parameters for each switch code are contained in Table 3 for reference. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 40 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPU T FREQ. H.S. SHIFT CAP. L.S. SHIFT CAP. INDICO 100® DSS PART NO. (See note 2) INDICO IQ® DSS PART NO. (See note 2) CMP 200® DSS PART NO. (See note 2) CGR (GE) Statorix 200A/240/260 (50/110 Ω stator) MN640 M641 MN641 M643 Type 28 (00111) LOW SPEED OPERATION ONLY (See note 1) 50 Hz N/A 12.5µF to 15.5µF 733317-15 735925-15 906894-15 (See note 10) 904925-23 901297-15 901298-15 (See note 10) 906672-23 MS740 MSN740 MSN742 RSN742 Type 28 (00111) 150 Hz 3 µF 12.5µF to 15.5µF 733317-15 735925-15 906894-15 (See note 10) 904925-23 901297-15 901298-15 (See note 10) 906672-23 RN620 RN621 Type 28 (00111) 50 Hz N/A 12.5µF to 15.5µF 733317-15 735925-15 906894-15 (See note 10) 904925-23 901297-15 901298-15 (See note 10) 906672-23 150 Hz 3 µF 12.5µF to 15.5µF CGR (GE) Statorix 550 (50/110 Ω stator) RSN722 Switches (1........5) Type 28 (00111) (SEE TABLE 3 FOR OPERATING PARAMETERS) Dual Speed Starters “-15” must be jumper-configured to provide the required 15.5 µF low speed shift capacitance. Please see note 10 before attempting to use this Dual Speed Starter configuration. HIGH SPEED OPERATION ONLY (See note 1) For Dual Speed operation, contact factory. LOW SPEED OPERATION ONLY (See note 1) Dual Speed Starters “-15” must be jumper-configured to provide the required 15.5 µF low speed shift capacitance. Please see note 10 before attempting to use this Dual Speed Starter configuration. HIGH SPEED OPERATION ONLY (See note 1) For Dual Speed operation, contact factory. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 41 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Chirana Rotax KA 125 (20/50 Ω stator, single phase) RIK-T Type 21 (10101) LOW SPEED OPERATION ONLY (See note 1) Comet DO7 / DX7 (25/50 Ω stator) DX7 DX71HS (See note 3) Type 0 (00000) Comet DO9 / DX9 (20/50 Ω stator) DX9 DX91H / HS DX92H / HS DX93H / HS DX94HS DX96HS DX97HS (See note 3) DX10H (See note 3) Comet DO10 / DX10 (20/50 Ω stator) DX10HS DX104HS (See note 3) DX105HS DX106HS (See note 3) 60 Hz H.S. SHIFT CAP. N/A L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 0 (00000) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 0 (00000) LOW SPEED OPERATION ONLY (See note 1) None 60 Hz N/A 30 µF to 36 µF 50/150 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 None 50/150 Hz 6 µF 30 µF to 36 µF Switches (1........5) Type 24 (00011) Type 24 (10011) (SEE TABLE 3 FOR OPERATING PARAMETERS) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Page 42 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, Comet DO700 / DX700 (25/50 Ω stator) DX700HS (See note 3) Dunlee (Philips) DA10 series (20/50 Ω stator) Dunlee (Philips) DA90 DA90HS (36/44 Ω stator) (See note 4) Dunlee (Philips) DA90 DA90HS (9/11 Ω stator) (See note 5) Dunlee (Philips) DR1400 (20/50 Ω stator) OPERATING NOTES OUTPUT FREQ. 50/150 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 24 (10011) None DA1002 DA1029 DA1036 DA1083 DA1092 DA1094 DU404 (0.6/1.25) DU 1750 DU 3050 DU 2550 DU 33100 (See note 6) Type 4 (00100) None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 9 (10010) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30µF DU 1750 DU 3050 (See note 6) DU 2550 (See note 6) DU 33100 (See note 6) Type 9 (10010) LOW SPEED OPERATION ONLY (See note 1) None 60 Hz N/A 37.5µF 60/180 Hz 12.5µF 37.5µF 180 Hz 12.5µF 37.5µF DR1429 DR1436 DR1492 DR1494 DU404 (0.6/1.25) Type 4 (00100) HIGH SPEED OPERATION ONLY (See note 1) None 60/180 Hz 6 µF 30 µF to 36 µF Switches (1........5) Type 9 (10010) Type 9 (10010) (SEE TABLE 3 FOR OPERATING PARAMETERS) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-23 901297-13 901298-13 906672-21 906672-23 733317-16 735925-16 906894-16 733317-17 735925-17 906894-17 904925-21 904925-23 901297-16 901298-16 906672-21 906672-23 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Rev. AF Page 43 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) OUTPUT FREQ. Dunlee (Philips) DR 1800 (19/22 Ω stator) (See note 4) DR1817 Type 9 (10010) LOW SPEED OPERATION ONLY (See note 1) Dunlee (Philips) DR 1800 (9/11 Ω stator) (See note 5) DR1817 Type 9 (10010) Type 9 (10010) Type 9 (10010) Type 7 (11100) DR1825 DR1833 Dunlee (Picker / Philips) PX1300 3” anode “S“ stator (15/30 Ω) PX1302 PX1312 PX1351 DU303 (1.0/2.0) Dunlee (Picker / Philips) PX1400 series 4” anode “S” stator (15/30 Ω) PX1402 PX1412 PX1415 PX1425 PX1429 PX1431 PX1436 PX1456 PX1463 PX1472 PX1473 PX1475 PX1482 PX1483 PX1492 PX1494 DU404 (0.6/1.25) PX1402 PX1412 PX1415 PX1425 PX1429 PX1431 PX1436 PX1456 PX1463 PX1472 PX1473 PX1482 PX1483 PX1492 PX1494 DU404 (0.6/1.25) Dunlee (Picker / Philips) PX1400 series 4” anode “Q” stator (6/12 Ω) 60 Hz H.S. SHIFT CAP. N/A L.S. SHIFT CAP. 30µF to 32µF INDICO 100® DSS PART NO. (See note 2) 733317-13 735925-13 906894-13 LOW SPEED OPERATION ONLY (See note 1) HIGH SPEED OPERATION ONLY (See note 1) None 60 Hz N/A 37.5µF 180 Hz 12.5µF 37.5µF 60/180 Hz 12.5µF 37.5µF 733317-16 735925-16 906894-16 733317-17 735925-17 906894-17 None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 1 (10000) None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 13 (10110) None 60/180 Hz 20 µF 60 µF to 62 µF 733317-02 735925-02 906894-02 Switches Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF INDICO IQ® DSS PART NO. (See note 2) 904556-13 903255-13 904925-21 904925-22 904925-23 904925-21 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-13 901298-13 906672-21 906672-22 906672-23 901297-16 901298-16 906672-21 906672-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 904925-22 901297-02 901298-02 906672-22 Page 44 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. H.S. SHIFT CAP. L.S. SHIFT CAP. INDICO 100® DSS PART NO. (See note 2) INDICO IQ® DSS PART NO. (See note 2) CMP 200® DSS PART NO. (See note 2) Type 14 (01110) None 60/180 Hz 7.5 µF 36 µF to 37.5µF 733317-12 735925-12 906894-12 904556-12 903255-12 904925-21 901297-12 901298-12 906672-21 Type 20 (00101) None 60/180 Hz 7.5 µF 36 µF to 37.5µF 733317-12 735925-12 906894-12 904556-12 903255-12 904925-21 901297-12 901298-12 906672-21 AR11-30 AR30-60 Type 22 (01101) LOW SPEED OPERATION ONLY (See note 1) 50 Hz N/A 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 AR20-50 AR30-100 AR40-100 Type 22 (01101) None 50/150 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Eureka (see Varex/Varian) GE Maxiray 75 (3” anode) 23/23Ω equal impedance “E” stator * Dunlee tube GE Maxiray 100 (4” anode) 23/23 Ω equal impedance “E” stator Gilardoni Rotagil (33/36 Ω stator) Gilardoni Rotagil A/A5 (33/36 Ω stator) MX-75 DU303 (1.0/2.0) * DU303 (0.6/1.2) * MX-100 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 45 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. 50 Hz H.S. SHIFT CAP. N/A L.S. SHIFT CAP. 30 µF to 36 µF IAE C30 C31 (25/62 Ω stator) X20P X22 X40 RTM70H RTM 70HS Type 10 (01010) LOW SPEED OPERATION ONLY (See note 1) Type 10 (01010) Additional boost time may be required at 400VAC (See note 12) LOW SPEED OPERATION ONLY (See note 1) 50/150 Hz 6 µF 60 Hz N/A 30 µF to 36 µF 20 µF to 30 µF IAE C40 (33/64 Ω stator) X39 X40 X42 Type 19 (11001) IAE C52 C52 Super C100 C352 (20/40 Ω stator) RTM 78H RTM 90H RTM 92H RTM 101H RTM 102H RTM 780H RTM 782H X39 X40 X42 X50H X50AH RTC 600HS RTC 700HS RTC 1000HS RTM 78HS RTM 90HS RTM 92HS RTM 101HS RTM 102HS RTM 780HS RTM 782HS Type 27 (11011) LOW SPEED OPERATION ONLY (See note 1) 50 Hz N/A 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Type 27 (11011) Additional boost time may be required at 400VAC (See note 12) 50/150 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 733317-17 735925-17 906894-17 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 Rev. AF Page 46 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Kailong RADII H2100 (20/50 Ω stator) KL100 Type 21 (10101) None Machlett (Varex) Dynamax 52 Std “R” stator (16/50 Ω) A-102 A-132 A-142 Type 0 (00000) Machlett (Varex) Dynamax 62 “STD” stator (23/56 Ω) Dynamax 62U configured as “STD” or “R” (15/36 Ω) A-192B A-196 A-197 Type 0 (00000) A-256 A-272 A-282 A-292 A-482 Type 1 (10000) A-286 Switches (1........5) 60/180Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 (SEE TABLE 3 FOR OPERATING PARAMETERS) INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 A-486 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 47 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Philips ROT 350 ROT 351 (36/44 Ω stator) (See note 4) RO 1230 RO 1750 RO 2050 RO 3050 SRO 2250 Type 9 (10010) LOW SPEED OPERATION ONLY (See note 1) Type 9 (10010) * Dunlee tube (See note 6) ** NAGO tube *** Varex (Varian) tube Philips ROT 350 ROT 351 (9/11 Ω stator) (See note 5) * Dunlee tube (See note 6) *** Varex (Varian) tube Philips ROT 500 ROT 501 (9/11 Ω stator) 60 Hz H.S. SHIFT CAP. N/A L.S. SHIFT CAP. 30µF to 32µF INDICO 100® DSS PART NO. (See note 2) 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-13 901298-13 906672-21 906672-22 906672-23 LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 37.5µF 904925-21 904925-23 901297-16 901298-16 906672-21 906672-23 Type 9 (10010) None 60/180 Hz 12.5µF 37.5µF 733317-16 735925-16 906894-16 733317-17 735925-17 906894-17 PG-256 *** PG-292 *** RAD-34 *** RAD-50 *** RO 2550 RO 33100 SRO 1330 DU 2550 * DU 33100 * SRM 0310 Type 9 (10010) HIGH SPEED OPERATION ONLY (See note 1) 180 Hz 12.5µF 37.5µF Type 30 (01111) None 60/180 Hz 12.5µF 37.5µF 904925-21 904925-23 SRM 1080 SRM 35100 Type 30 (01111) HIGH SPEED OPERATION ONLY (See note 1) 180 Hz 733317-16 735925-16 906894-16 733317-17 735925-17 906894-17 901297-16 901298-16 906672-21 906672-23 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) DU 1750 * DU 3050 * GD6 3050 ** PG-256 *** PG-292 *** RAD-34 *** RAD-50 *** RO 1230 RO 1750 RO 3050 DU 1750 * DU 3050 * RO 2050 SRO 2250 37.5µF Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 48 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. H.S. SHIFT CAP. L.S. SHIFT CAP. INDICO 100® DSS PART NO. (See note 2) INDICO IQ® DSS PART NO. (See note 2) CMP 200® DSS PART NO. (See note 2) Shimadzu (Circlex) RX80 RX81 RX82 (7.5/35 Ω stator) P18DE-85 Type 16 (00001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF 733317-13 735925-13 906894-13 P33D P38D P38DE-85 Type 16 (00001) HIGH SPEED OPERATION ONLY (See note 1) 180 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 Shimadzu (Circlex) RX100 P38E Type 16 (00001) None 60/180 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 Shimadzu (Circlex) RX150 (8.4/24 Ω stator) P18DK Type 16 (00001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF 733317-13 735925-13 906894-13 P38DK P324DK Type 16 (00001) HIGH SPEED OPERATION ONLY (See note 1) 180 Hz 5 µF 30 µF 904556-13 903255-13 904925-21 904925-22 904556-13 903255-13 904925-21 904925-22 904556-13 903255-13 904925-21 904925-22 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 901297-13 901298-13 906672-21 906672-22 901297-13 901298-13 906672-21 906672-22 901297-13 901298-13 906672-21 906672-22 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Picker (see Dunlee) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 49 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, Siemens Biangulix “S” stator (1-phase only) (14/18 Ω) Siemens Opti150 “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube Siemens OptiTop “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube Siemens Optilix “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube Siemens Megalix Siemens SV 125 “S” stator (1-phase only) (14/18 Ω) BI 125/20/40 BI 125/30/50 BI 150/30/50 BI 150/30/52R 150/30/52R 150/40/72C 150/12/50 150/40/80 150/30/50 150/40/102C SG-796B * 150/40/80HC-100L 150/40/80HC-102L OPERATING NOTES OUTPUT FREQ. Type 23 (11101) HIGH SPEED OPERATION ONLY (See note 1) Type 23 (11101) HIGH SPEED OPERATION ONLY (See note 1) Switches (1........5) Type 23 (11101) L.S. SHIFT CAP. 30 µF INDICO 100® DSS PART NO. (See note 2) 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-13 903255-13 904925-21 904925-22 CMP 200® DSS PART NO. (See note 2) 901297-13 901298-13 906672-21 906672-22 150 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 HIGH SPEED OPERATION ONLY (See note 1) 150 Hz 5 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 30 µF Type 23 (11101) HIGH SPEED OPERATION ONLY (See note 1) 150 Hz 5 µF HIGH SPEED OPERATION ONLY (See note 1) 30 µF 30 µF None Type 8 (00010) 30 µF 30 µF None SG-256 * 125/30/82CM-120LW 125/40/82CM-120LW 150 Hz H.S. SHIFT CAP. 5 µF None SG-1096 * 150/30/50C-100L 150/40/80C-100L (SEE TABLE 3 FOR OPERATING PARAMETERS) 150 Hz 6 µF 30 µF Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 50 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Siemens SV 150 “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube SV 150/30/50C100L Type 23 (11101) HIGH SPEED OPERATION ONLY (See note 1) 150 Hz Siemens SV 150 “S” stator (1-phase only) (14/18 Ω) * Varex (Varian) tube SV 150/40/80C100L Siemens (SXVT) RAY-6_1 RAY-8(S)_1 RAY-12(S)_1 (1-phase only) (20/50 Ω stator) Switches (1........5) SG-296B * (SEE TABLE 3 FOR OPERATING PARAMETERS) H.S. SHIFT CAP. 5 µF L.S. SHIFT CAP. 30 µF INDICO 100® DSS PART NO. (See note 2) 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-13 903255-13 904925-21 904925-22 CMP 200® DSS PART NO. (See note 2) 901297-13 901298-13 906672-21 906672-22 30 µF None Type 23 (11101) None 60/180 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 SV 150/12/40-S SV 150/25/48-S SV 150/22/54-S Type 31 (11111) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 904925-23 901297-13 901298-13 906672-21 906672-22 906672-23 Siemens (SXVT) RAY-13(S)_1 RAY-14(S)_1 (1-phase only) (20/50 Ω stator) SV 150/33/78R-S Type 31 (11111) None 60/180 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 Siemens (SXVT) SDR (1-phase only) (20/50 Ω stator) SDR 150/30/50-1 Type 31 (11111) None 60/180 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 Siemens (SXVT) SV 150 “S” stator (1-phase only) (14/18 Ω) SV 150/40/80C100_1 Type 23 (11101) None 150 Hz 5 µF 30 µF 733317-13 735925-13 906894-13 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 SG-796B * (See note 7) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 51 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) Toshiba Rotanode Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing used. For ease of use, the Toshiba tubes are sorted by Tube Type (“E” number) within this table. The installer must confirm the stator type for the planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the appropriate table entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same line within this table. DRX-3624H 30 µF to 733317-12 Type 17 None 60/180 Hz 6 µF 904556-12 901297-12 DRX-3724H 735925-12 903255-12 901298-12 36 µF (10001) DRX-4634H 906894-12 901297-13 904556-13 901298-13 733317-13 903255-13 735925-13 906672-21 904925-21 904925-22 906894-13 906672-22 904925-23 906672-23 30 µF to 733317-12 E7100 Type 17 None 60/180 Hz 6 µF 901297-12 904556-12 (See note 8) 735925-12 901298-12 903255-12 36 µF (10001) 906894-12 904556-13 901297-13 733317-13 903255-13 901298-13 735925-13 904925-21 906672-21 906894-13 904925-22 906672-22 904925-23 906672-23 20 µF to LOW SPEED E7132 Type 19 60 Hz N/A 733317-13 904556-13 901297-13 OPERATION ONLY E7239 735925-13 903255-13 901298-13 30 µF (11001) (See note 1) E7240 906894-13 904925-21 906672-21 904925-22 E7242 906672-22 (See note 8) 904925-23 * 906672-23 * * Preferred * Preferred Toshiba XH-112V-2 XH-112W XS-AA stator XS-AG stator (9.4/28.3 Ω) Toshiba XH-112V XS-AG stator (9.4/28.3 Ω) Toshiba XH-121 XH-126 XH-150 XS-AV stator XS-RA stator (27.5/58 Ω) Toshiba XH-112V XS-AG stator (9.4/28.3 Ω) E7250 (See note 8) CODE, Switches (1........5) Type 17 (10001) OPERATING NOTES (SEE TABLE 3 FOR OPERATING PARAMETERS) None OUTPUT FREQ. 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Rev. AF INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Page 52 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Toshiba XH-106V XH-180 XH-181 XS-AL stator (9.4/28.3 Ω) E7252 (See note 8) Type 29 (10111) None Toshiba XH-106V XH-180 XH-181 XS-RA stator (27.5/58 Ω) E7252 (See note 8) Type 19 (11001) Switches (1........5) 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Dual Speed Starters “-15” must be jumper-configured to provide the required 28 µF low speed shift capacitance. Please refer to note 10 before attempting to use this Dual Speed Starter configuration. LOW SPEED OPERATION ONLY (See note 1) 60/180 Hz 3 µF 23 µF to 28 µF 733317-15 735925-15 906894-15 (See note 10) 60 Hz N/A 20 µF to 30 µF 733317-13 735925-13 906894-13 (SEE TABLE 3 FOR OPERATING PARAMETERS) Toshiba XH-157 XS-RB stator (20.2/38 Ω) E7254 E7255 (See note 8) Type 21 (10101) None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Toshiba XH-121 XH-126 XS-AV stator XS-RA stator (27.5/58 Ω) E7299 (See note 8) Type 19 (11001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 20 µF to 30 µF 733317-13 735925-13 906894-13 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-15 901298-15 (See note 10) 906672-23 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred Page 53 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Toshiba XH-106V XH-180 XS-AL stator (9.4/28.3 Ω) E7813 (See note 8) Type 29 (10111) None Toshiba XH-121 XS-BA stator (18/47.5 Ω) E7843 (See note 8) Type 0 (00000) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Toshiba XH-121 XH-126 XH-150 XS-RA stator (27.5/58 Ω) E7861 (See note 8) Type 19 (11001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 20 µF to 30 µF 733317-13 735925-13 906894-13 Toshiba XH-112V XS-AG stator (9.4/28.3 Ω) E7864 (See note 8) Type 17 (10001) None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Toshiba XH-121 XH-126 XS-AV stator (27.5/58 Ω) E7865 (See note 8) Type 19 (11001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 20 µF to 30 µF 733317-13 735925-13 906894-13 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred Page 54 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Toshiba XH-112V XS-AG stator (9.4/28.3 Ω) E7869 (See note 8) Type 17 (10001) None Toshiba XH-121 XS-AV stator XS-RA stator (27.5/58 Ω) E7876 (See note 8) Type 19 (11001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 20 µF to 30 µF 733317-13 735925-13 906894-13 Toshiba XH-121 XH-126 XH-150 XS-AL stator (9.4/28.3 Ω) E7884 (See note 8) Type 29 (10111) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 Toshiba XH-121 XS-126 XS-AV stator XS-RA stator (27.5/58 Ω) E7886 (See note 8) Type 19 (11001) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 20 µF to 30 µF 733317-13 735925-13 906894-13 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 904556-13 903255-13 904925-21 904925-22 904925-23 * * Preferred CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 901297-13 901298-13 906672-21 906672-22 906672-23 * * Preferred Page 55 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, Varex (Varian) B-100 “R” stator (16/50 Ω) A-102 A-132 / A-134* A-142 / A-144* * (See note 9) A-145 A-102 A-132 / A-134* A-142 / A-144* * (See note 9) Type 11 (11010) Varex (Varian) B-100 “Q” stator (8/15 Ω) OPERATING NOTES OUTPUT FREQ. Type 0 (00000) None Type 2 (01000) LOW SPEED OPERATION ONLY (See note 1) None Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 60 Hz N/A 60/180 Hz 20 µF 60 µF to 62 µF 733317-02 735925-02 906894-02 904925-22 901297-02 901298-02 906672-22 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 56 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Varex (Varian) B-130 B-130H B-135H B-146 B-147 B-150 B-150H “R” stator (16/50 Ω) A-152 A-182 / A-184* A-192 / A-194* A-195 A-196 A-197 * (See note 9) A-272 / A-274* A-277 / A-278* A-282 / A-284 * A-286 A-292 / A-294* A-482 G-242 G-256 G-292 / G-294* G-296 G-892 G-896 * (See note 9) A-152 A-182 / A-184* A-192 / A-194* A-195 / A-196 A-197 / A-252 A-272 / A-274* A-277 / A-278* A-282 / A-284 * A-286 A-292 / A-294* A-482 G-242 G-256 G-292 / G-294* G-296 RAD-60 * (See note 9) Type 0 (00000) None None 60/180 Hz 20 µF 60 µF to 62 µF 733317-02 735925-02 906894-02 904925-22 901297-02 901298-02 906672-22 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Varex (Varian) B-130 B-130H B-135H B-150 B-150H “Q” stator (6/11 Ω) Switches (1….....5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Type 1 (10000) Type 12 (00110) Page 57 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Varex (Varian) B-160 B-160H B-165H “R” Stator (16/50 Ω) G-1077 G-1078 G-1080 G-1082 / G1084* G-1086 / G1087* G-1092 / G-1094* * (See note 9) G-1582 G-1592 G-1593 Type 15 (11110) None Type 15 (11110) SG-1096 Varex (Varian) B-240H “R” Stator (14/46 Ω) Varex (Varian / Eureka) Diamond Emerald LEO “R” stator (20/50 Ω) Varex (Varian) B-180 B-180H B-185 B-185H “R” Stator (16/50 Ω) Varex (Varian) B-199 “S” Stator (15/18 Ω) 60/180 Hz H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Type 18 (01001) None 50/150 Hz 5 µF 30 µF 904556-13 903255-13 904925-21 904925-22 901297-13 901298-13 906672-21 906672-22 G-2090 Type 6 (01100) LIMITED USAGE – CONTACT FACTORY 100 / 180 Hz 6 µF 12.5µF to 20µF 733317-13 735925-13 906894-13 733317-17 735925-17 906894-17 733317-16 735925-16 906894-16 904925-21 904925-22 904925-23 RAD-8 RAD-12 RAD-74 Type 0 (00000) LOW SPEED OPERATION ONLY (See note 1) 60 Hz N/A 30 µF to 36 µF RAD-13 RAD-14 RAD-68 Type 0 (00000) None 60/180 Hz 6 µF 30 µF to 36 µF 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 901297-16 901298-16 906672-21 906672-22 906672-23 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 58 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued) TUBE TYPE (HOUSING) TUBE TYPE (INSERT) CODE, OPERATING NOTES OUTPUT FREQ. Varex (Varian / Eureka) Sapphire “R” stator (20/50 Ω) RAD-21 RAD-40 RAD-44 RAD-56 RAD-60 * RAD-92 RAD-94 Type 5 (10100) None 60/180 Hz * For “Q” stator, refer to B-130 housing Switches (1........5) (SEE TABLE 3 FOR OPERATING PARAMETERS) H.S. SHIFT CAP. 6 µF L.S. SHIFT CAP. 30 µF to 36 µF INDICO 100® DSS PART NO. (See note 2) 733317-12 735925-12 906894-12 733317-13 735925-13 906894-13 INDICO IQ® DSS PART NO. (See note 2) 904556-12 903255-12 904556-13 903255-13 904925-21 904925-22 904925-23 CMP 200® DSS PART NO. (See note 2) 901297-12 901298-12 901297-13 901298-13 906672-21 906672-22 906672-23 Note for Table 2 NOTE 1: Tube types designated as low speed only or high speed only must be programmed for low speed only or high speed only operation. Refer to the section Tube Selection within Chapter 3 of the applicable generator service manual for details. For some tubes listed as high speed only, the starter may be capable of low speed operation but the Manufacturer’s data sheet for the Insert lists high speed only. In other cases, the limitation is due to operational limitations of the Dual Speed Starter itself. NOTE 2: Dual Speed Starter part numbers 733317-XX (Indico 100®), 904556-XX (Indico IQ®) and 901297-XX (CMP 200® DR) are used in 400 VAC 3φ generators / power supplies and in 480 VAC 3φ generators / power supplies with a line adjusting transformer. Dual Speed Starter part numbers 735925-XX (Indico 100®), 903255-XX (Indico IQ®) and 901298-XX (CMP 200® DR) are used in 230 VAC 1φ generators / power supplies Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 59 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc and in 480 VAC 3φ direct input generators / power supplies (using no line adjusting transformer). Dual Speed Starter part numbers 906894-XX (Indico 100®), 904925-XX (Indico IQ®) and 906672-XX (CMP 200® DR) are used in both 400 VAC and 480 VAC 3φ generators. These RoHS-compliant Starters replace the existing 73317-XX / 735925-XX series (Indico 100®), 904556-XX / 903255-XX series (Indico IQ®) and 901297-XX / 901298-XX series (CMP 200® DR), and may provide a greater selection of available shift capacitance. Where more than one Dual Speed Starter part number is referenced for a particular tube type, any of these (applicable to the generator and input voltage) is acceptable. For 2-tube systems using different tube types, both tube types must be considered. Where available, the XXXXXX-21 starter is generally preferred, unless otherwise indicated in the tables. The Dual Speed Starters provide a selection of shift capacitance values for both high speed and low speed operation, as identified by the DSS part number suffix (“tab type”). Unless otherwise specified, capacitor selection is made automatically by relays located on the Dual Speed Starter board. The following table lists the capacitance values provided, along with applicable notes. Also, refer to TABLE: DSS BOARD AND SHIFT CAPACITANCE CONFIGURATIONS Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 60 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc TABLE: DSS BOARD AND SHIFT CAPACITANCE CONFIGURATIONS (NOTE 2 Continued) DSS Part No. (applicability) HIGH-SPEED SHIFT CAPACITANCE LOW-SPEED SHIFT CAPACITANCE COMMENTS XXXXXX-01 (INDICO 100® only) XXXXXX-02 (All except Indico IQ®) XXXXXX-12 (All except Indico IQ®) XXXXXX-13 (All except Indico IQ®) XXXXXX-15 (All except Indico IQ®) 6 µF 31 µF May be replaced by XXXXXX-21 in all configurations where available. 20 µF 40 µF 6 µF, 7.5 µF 36 µF, 37.5 µF Typically used to support “Q” stator tubes. May be replaced by XXXXXX-22 in all configurations, where available. May be replaced by XXXXXX-21 in all configurations, where available. 5 µF, 6 µF 30 µF May be replaced by XXXXXX-21 in all configurations, where available. 3 µF, 6 µF Typically used to support CGR (GE) Statorix or Toshiba E7252 (XS-R/RA) tubes. May be replaced by XXXXXX-23 in all configurations, where available. Manual configuration required; see Note 10. XXXXXX-16 (All except Indico IQ®) 6 µF, 12.5 µF 15.5 µF, 18.5 µF or 28 µF, 31 µF (manually selected) 12.5 µF, 31.5 µF, 37.5 µF XXXXXX-17 (INDICO 100® only) 5 µF, 12.5 µF 30 µF, 37.5 µF XXXXXX-21 (All except Indico 100®) 5 µF, 6 µF, 7.5 µF, 12.5 µF 5 µF, 6 µF, 20 µF 3 µF, 6 µF, 12.5 µF 12.5 µF, 30 µF, 37.5 µF 20 µF, 30 µF, 62.5 µF 23 µF, 32.5 µF XXXXXX-22 (All except Indico 100®) XXXXXX-23 (All except Indico 100®) Used to support Philips ROT350/500 tubes. May be used in place of Dual Speed Starters “-12” or “-13” in two-tube systems that include both a standard “R” stator tube (6 µF high speed shift capacitance) and a Philips ROT 350/500 tube. All such configurations may be replaced by XXXXXX-21, where available. Used in place of Dual Speed Starter “-13” in two-tube systems that include both a standard “R” stator tube (6 µF of high speed shift capacitance) and an IAE tube (5 µF of high speed shift capacitance). All such configurations may be replaced by XXXXXX-21, where available. Do not use these starters in place of XXXXXX-13 with Siemens tubes due to high-speed shift capacitor voltage limitations. Generally preferred where available, unless support may be required for Toshiba E7252 tubes (see XXXXXX-23), unless otherwise indicated within the tables. Typically used to support “Q” stator tubes. Preferred over XXXXXX-02, where available. Typically used where support may be required for CGR (GE) Statorix or Toshiba E7252 (XS-R/RA) tubes. Preferred over XXXXXX-15, where available. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 61 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc NOTE 3: Comet tube inserts with the prefix “DI” and “DX” are interchangeable. NOTE 4: Philips / Dunlee housing with the windings connected in series (low-speed / high impedance configuration). NOTE 5: Philips / Dunlee housing with the windings connected in parallel (highspeed / low impedance configuration). NOTE 6: Dunlee tube inserts with the prefix “DU” and Philips tube inserts with the prefix “RO” are interchangeable. Select the corresponding Philips insert type within the Generator software if the Dunlee type is not available. NOTE 7: Siemens (SXVT) housing with the “phase-shifted” windings connected in parallel (low impedance, high speed configuration). NOTE 8: Complete Toshiba x-ray tube assemblies include the suffix “X, “FX”, “GX”, or “JX”, which are electrically interchangeable. NOTE 9: These X-ray tubes incorporate a control grid. Grid control is currently not supported by CPI generators. Connect the grid connection to Cathode Common when using these tubes, and select the insert type within the Generator software corresponding to the equivalent non-grid tube. NOTE 10: Dual Speed Starters 733317-15 / 735925-15 / 906894-15 (Indico 100) and 901297-15 / 901298-15 (CMP 200® DR) have jumper selectable phase shift capacitance that allows selection of 15.5 µF or 28 µF low speed shift capacitors for operation with CGR (GE) Statorix or Toshiba E7252 (XS-R/RA stator) tubes, respectively, as per table 2. By default, this starter is factory set to the 28 µF position. Confirm proper configuration per the section “Configuring Dual Speed Starter XXXXXX-15” in chapter 2 before proceeding. These starters also support standard “R” stators that require 6 µF of high speed shift capacitance. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 62 Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators CPI Canada Inc (INDICO 100® only): If configured for 28 µF of low speed capacitance, these Dual Speed Starters also support tubes listed in Table 2 as requiring Dual Speed Starters 733317-13 / 735925-13 / 906894-15 and requiring 6 µF of high speed capacitance. For two-tube operation, this configuration will support: 1. These tubes and the Toshiba E7252/XS-RA (high and low speed) simultaneously, or; 2. These tubes and a CGR Statorix tube (high-speed only) simultaneously 3. The Toshiba E7252/XS-RA and a CGR tube simultaneously if one or both of these tubes is operated in high-speed mode only. (CMP 200® DR only): If configured for 28 µF of low speed capacitance, these Dual Speed Starters also support tubes listed in Table 2 as requiring Dual Speed Starters 901297-13 / 901298-13 and requiring 6 µF of high-speed capacitance. NOTE 11: When operating under 400 VAC low-line conditions, tubes requiring a High Speed start voltage of 400 V or greater (refer to Table 3) may not be provided with full power high-speed boost. In this case, the boost time may need to be extended and the anode speed must be checked. NOTE 12: The Tube Type (Code) indicated for this tube requires greater than 400 VAC line input for full power high-speed boost. When used with 400 VAC line, the boost time may need to be extended and the anode speed must be checked. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 63 Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes CPI Canada Inc (This page intentionally left blank) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 64 Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes CPI Canada Inc Dual-Speed Starter and Indico IQ Low-Speed Starter X-ray Tube Codes Tube operating parameters for each tube type are selected by setting switches to the applicable Code, as indicated in Table 1C (Indico IQ LSS) and Table 2 (DSS). For reference, operating parameters applicable to each Code are contained in Table 3 below. Reference notes are provided at the end of Table 3. Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 65 Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes CPI Canada Inc TABLE 3: TUBE TYPE CODES (DUAL SPEED STARTER and INDICO IQ® LOW SPEED STARTER) TUBE TYPE (CODE) CODE Switches (1........5) H.S. START VOLTS H.S. RUN VOLTS H.S. BOOST TIME H.S. FREQ. H.S. BRAKE VOLTS H.S. BRAKE TIME L.S. START VOLTS L.S. RUN VOLTS L.S. BOOST TIME L.S. FREQ. Hz 60 Hz H.S. SHIFT CAP. 6 µF 0 00000 400 V 100 V 1.4 sec 180 Hz 100 V 3.0 sec 240 V 60 V 1.4 sec 1 10000 400 V 100 V 1.9 sec 180 Hz 100 V 3.0 sec 240 V 70 V 2 01000 400 V 100 V 1.0 sec 180 Hz 100 V 3.0 sec 240 V 60 V 3 11000 400 V 100 V 5.0 sec* 180 Hz 150 V 3.0 sec 240 V 4 00100 420 V 70V 2.5 sec 180 Hz 170 V 3.0 sec* 5 10100 400 V 100 V 2.3 sec 180 Hz 100 V 6 01100 440 V 140 V 6.0 sec* 180 Hz 7 11100 240 V 120 V 2.3 sec 8 00010 400 V 90 V 9 10010 280 V 10 01010 11 -12, -13, -15, -16, -21, -22, -23 1.9 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 1.0 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 70 V 5.0 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 * ( 2.5 sec Ready) 240 V 50 V 1.6 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 * (LS braking 0.8 sec) 3.0 sec 240 V 50 V 2.3 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 250 V 4.0 sec** 400 V 110 V 5.0 sec 100 Hz 6 µF *** -16, -21, -22, -23 * ( 5.0 sec Ready) ** (LS braking 2.0s) *** (12.5µF-20µF L.S. cap.) 180 Hz 100 V 3.0 sec 240 V 50 V 2.3 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 3.0 sec 150 Hz 100 V 3.0 sec 240 V 100 V 1.5 sec 50 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 60 V 1.5 sec 180 Hz 80 V 3.0 sec 180 V 50 V 1.5 sec 60 Hz 12.5 µF -12 (LS only), -16, -17, -21, -22 (LS only), -23 440 V 100 V 0.8 sec 150 Hz 120 V 1.8 sec 220 V 50 V 0.6 sec 50 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 11010 290 V 70 V 0.8 sec 180 Hz 60 V 3.0 sec 150 V 50 V 0.8 sec 60 Hz 20 µF -02, -22 12 00110 290 V 60/80V 1.3 sec 180 Hz 60 V 3.0 sec 150 / 180 V 50 V 1.3 sec 60 Hz 20 µF -02, -22 13 10110 340 V 60 V 1.0 sec 180 Hz 100 V 3.0 sec 240 V 70 V 1.0 sec 60 Hz 20 µF -02, -22 14 01110 400 V 90 V 0.9 sec 180 Hz 80 V 2.0 sec 230 V 70 V 0.9 sec 60 Hz 7.5 µF -12, -21 15 11110 400 V 100 V 5.0 sec 180 Hz 150 V 3.0 sec 240 V 70 V 5.0 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 16 00001 340 V 80 V 1.6 sec 180 Hz 80 V 3.0 sec 150 V 50 V 1.6 sec 60 Hz 5 µF -13, -17, -21, -22, -23 (LS only) 17 10001 320 V 80 V 2.3 sec 180 Hz 100 V 3.0 sec 200 V 50 V 1.5 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 18 01001 400 V 100 V 2.8 sec 150 Hz 120 V 3.0 sec 180 V 60 V 1.5 sec 50 Hz 5 µF -13, -17, -21, -22, -23 (LS only) 19 11001 320 V 130 V 2.0 sec 180 Hz 130 V 3.0 sec 240 V 60 V 1.4 sec 60 Hz 3 µF -12, -13, -16, -21,-22 (all LS only), -15, -23 (LS/HS) 20 00101 400 V 90 V 1.0 sec 180 Hz 80 V 3.0 sec 230 V 70 V 1.0 sec 60 Hz 7.5 µF -12, -21 Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF APPLICABLE DSS TYPES (See note 1) Page 66 Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes CPI Canada Inc TABLE 3: TUBE TYPE CODES (DUAL SPEED STARTER and INDICO IQ® LOW SPEED STARTER) TUBE TYPE (CODE) CODE Switches (1........5) H.S. START VOLTS H.S. RUN VOLTS H.S. BOOST TIME H.S. FREQ. H.S. BRAKE VOLTS H.S. BRAKE TIME L.S. START VOLTS L.S. RUN VOLTS L.S. BOOST TIME L.S. FREQ. Hz APPLICABLE DSS TYPES (See note 1) 60 Hz H.S. SHIFT CAP. 6 µF 21 10101 420 V 100 V 2.2 sec 180 Hz 150 V 3.0 sec 240 V 80 V 1.8 sec 22 01101 340 V 60 V 1.4 sec 150 Hz 120 V 4.0 sec 220 V 60 V 23 11101 360 V 90 V 1.6 sec 150 Hz 80 V 3.0 sec 150 V 50 V 1.5 sec 50 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 1.6 sec 50 Hz 5 µF -13, -21, -22, -23 (LS only) 24 00011 420 V 80 V 1.8 sec 150 Hz 150 V 3.0 sec 240 V 80 V 1.8 sec 50 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 25 10011 420 V 80 V 2.2 sec 150 Hz 150 V 3.0 sec 240 V 80 V 1.8 sec 50 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 26 01011 400 V 80 V 1.0 sec 150 Hz 120 V 27 11011 440 V 100 V 1.1 sec 150 Hz 120 V 3.0 sec 240 V 80 V 1.0 sec 50 Hz 5 µF Reserved 1.8 sec 220 V 80 V 1.1 sec 50 Hz 5 µF -13, -17, -21, -22, -23 (LS only) 28 00111 450 V 150 V 3.0 sec 150 Hz 120 V 3.0 sec 240 V 90 V 3.0 sec 50 Hz 3 µF -15, -21 (LS only), -23 29 10111 240 V 60 V 2.2 sec 180 Hz 80 V 1.6 sec 130 V 50 V 1.3 sec 60 Hz 6 µF -12, -13, -15, -16, -21, -22, -23 30 01111 420 V 50 V 1.8 sec 180 Hz 150 V 3.0 sec* 240 V 50 V 1.8 sec 60 Hz 12.5 µF -16, -17, -21, -23 * (LS braking: 1.2 sec) 31 11111 330V 80 V 1.4 sec 180 Hz 70V 3.0 sec 240 V 80 V 1.4 sec 60 Hz 5 µF -13, -21, -22, -23 (LS only) Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF -12, -13, -15, -16, -21, -22, -23 Page 67 Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes CPI Canada Inc Notes and Warnings for Table 3 Warning: The starter uses modulation strategies to obtain the desired outputs. Measured voltages may not agree with those listed in the table. However, the currents flowing in the stator windings are equivalent to those that would exist if the stator was excited with these voltages. NOTE 1: Tube Type codes should only be used with the applicable DSS types specified in this table (refer to Table 1 for applicable DSS Types). Refer to Table 1 for Indico IQ LSS Part Numbers applicable to each Tube Type (Housing and Insert). Refer to Table 2 for DSS Part Numbers applicable to each Tube Type (Housing and Insert). Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document. X-ray Tube Stator Compatibility Tables P/N: 746026-00 Rev. AF Page 68