X-ray Tube Stator
Compatibility Tables
Supplement for CPI X-ray Generators
P/N: 746026-00 Revision: AF
CPI Canada Inc.
Proprietary and Confidential
The information contained in this document are proprietary to
Communications & Power Industries Canada Inc. (CPI), have been
designed and developed at private expense, and are the exclusive
property of CPI. The information contained herein is loaned in
confidence and may not be duplicated in whole or in part, or be
used for re-engineering, reverse engineering, or otherwise
reproducing in any form or creating or attempting to create or
permitting, allowing or assisting others to create or manufacture
CPI’s product or products derived therefrom, without the express
written permission of CPI.
Copyright © By Communications & Power
Industries Canada Inc. All Rights Reserved.
Table of Contents
CPI Canada Inc
Contents:
Introduction ................................................................................ 3
Low-Speed Starter Tube Select Table ............................................. 4
Notes and Warnings ...................................................................................... 4
CMP 200® Generators ................................................................................ 4
CMP 150 ™ Generators ............................................................................... 4
Mobile DR Generators ................................................................................. 5
Indico IQ® Generators ............................................................................... 5
Indico 100® Generators ............................................................................. 6
Low-Speed Starter Tube Select Table – CMP 200® / CMP 150™ Generators ......... 7
Low-Speed Starter Tube Select Table – Mobile DR Generators .......................... 19
Low-Speed Starter Tube – Indico IQ® / Indico 100® Generators ...................... 23
Notes for Tables 1A, 1B, and 1C .................................................................... 38
Dual-Speed Starter Tube Select Table ........................................... 39
Note for Table 2 .......................................................................................... 59
Dual-Speed Starter and Indico IQ Low-Speed Starter X-ray Tube
Codes ....................................................................................... 65
Notes and Warnings for Table 3 .................................................................... 68
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 1
Table of Contents
CPI Canada Inc
(This page intentionally left blank)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 2
Introduction
CPI Canada Inc
Introduction
This supplement contains the X-Ray Tube Stator Compatibility Tables for the Low Speed Starter (LSS) and the Dual Speed Starter
(DSS). It is recommended to refer to the following Supplement Guide for the proper X-ray tube service.
Table 1A: Low-Speed Starter X-ray Tube Type
Selection for CMP 150 / CMP 200 Generator
and Notes and Warning
Table 1B: Low-Speed Starter X-ray Tube Type
Selection for Mobile DR Generator and Notes and
Warning
Table 1C: Low-Speed Starter X-ray Tube Type
Selection for Indico IQ / Indico 100 Generator
and Notes and Warning
Table 2: Dual-Speed Starter X-ray Tube Type
Selection for Indico IQ / Indico 100 / CMP 200
Generator and Notes for Table 2
Table 3: Dual-Speed Starter and Indico IQ LowSpeed Starter X-ray Tube Codes and Notes and
Warnings
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 3
Low-Speed Starter X-ray Tube Select Table
CPI Canada Inc
Low-Speed Starter Tube Select Table
This section contains notes and warnings for using low speed starter tube select
table provided in this supplement.
Notes and Warnings
CMP 200® Generators
The boost voltage may be set to either 240 VAC (factory default) or 120 VAC. This
should be set to 240 VAC, except where specifically noted in Table 1A. The Low
Speed Starter output is at the same frequency as the AC line (50 Hz or 60 Hz).
The Low Speed Starter is integrated into the HV Auxiliary Board within the CMP
200® and CMP 200® DR generators. The phase-shift capacitor is chassis-mounted
and may be changed in the field. The LSS part number corresponds to the value of
capacitor installed. The boost time is software-configurable from the console or
GenWare®. Note that these generators do not use a separate run voltage, instead
cycling the boost voltage on/off as required. The boost duty cycle is not hardwareconfigurable.
CMP 150 ™ Generators
Warning: The boost voltage is the same as the AC line voltage and is not
adjustable. Most X-ray tubes require a nominal 200 – 240 VAC supply. A
few X-ray tubes are intended to be operated using a 100 – 120 VAC
supply, as indicated within the tables.
In some cases, operating at tube intended for 200-240 VAC at 100-120 VAC may be
feasible, but requires increasing the boost time. If the alternate boost time is not
indicated within the tables, contact the factory for assistance.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 4
Low-Speed Starter X-ray Tube Select Table
CPI Canada Inc
The boost time is software-configurable from the console or GenWare®. Note that
these generators do not use a separate run voltage, instead cycling the boost
voltage on/off as required. The boost duty cycle is not hardware-configurable.
The Low Speed Starter output is at the same frequency as the AC line (50 Hz or
60 Hz).
The Low Speed Starter is integrated into the Input-Multifunction-Room (IMR)
Board within the CMP 150 generators. The phase-shift capacitor is chassismounted and may be changed in the field. The LSS part number corresponds to
the value of capacitor installed.
Mobile DR Generators
The Low Speed Starter is a separate sub-assembly within Mobile DR generator
configurations. The boost voltage is 240 VAC at 60 Hz, boost time is 1.5 seconds,
and a 30 µF phase-shift capacitor is provided.
Indico IQ® Generators
The Low-Speed-Starter (LSS) synthesizes its output frequencies independently of
the line frequency. It will operate most tubes at 60 Hz, as indicated in Table 1C.
Unless indicated otherwise within Table 1C, setting the switches to the code
indicated for the applicable Tube Type selects all required operating parameters.
The Low Speed Starter part number corresponds to the configuration of phaseshift capacitors installed.
Reference notes are provided at the end of Table 1C. Tube operating parameters
for each switch code are contained in Table 3 for reference.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 5
Low-Speed Starter X-ray Tube Select Table
CPI Canada Inc
Indico 100® Generators
The boost voltage may be set to either 240 VAC (factory default) or 120 VAC. This
should be set to 240 VAC, except where specifically noted in Table 1C. The Low
Speed Starter output is at the same frequency as the AC line (50 Hz or 60 Hz).
The Low Speed Starter is a separate sub-assembly within the INDICO
100®Generator. The LSS part number corresponds to the value of phase-shift
capacitor incorporated within this sub-assembly. The boost time is selectable to
either 1.5 seconds or 2.5 seconds, and the run voltage is selectable to any of 52
VAC, 73 VAC or 94 VAC.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 6
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
Low-Speed Starter Tube Select Table – CMP 200® / CMP 150™ Generators
Low-Speed-Starter X-ray Tube Types
Table 1A: CMP 200® / CMP 150™ Generators
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 7
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
240 VAC 1.5 sec 30 µF
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Chirana
Rotax KA 125
(20/50 Ω stator)
RIK-T 0.8/2.0 12/50
RIK-T 1.2/2.0 30/50
Comet
DO7 / DX7
(25/50 Ω stator)
DX7
DX71HS
(See note 3)
240 VAC
Comet
DO9 / DX9
(20/50 Ω stator)
DX9
DX91H / HS
DX92H / HS
DX93H / HS
DX94HS
DX96HS
DX97HS
(See note 3)
DX10H / HS
DX101H / HS
DX104HS
DX105HS
DX106HS
(See note 3)
XSTAR8 (XST-8)
XSTAR74 (XST-74)
XSTAR-WHIS
240 VAC
DA1002 (DU304-1.0/2.0)
DA1029 (DU304-0.6/1.2)
DA1036 (DU404-0.6/1.2)
DA1083 (DU404-0.3/1.0)
DA1092 (DU604-0.6/1.2)
DA1094 (DU404-0.4/0.8)
240 VAC
Comet
D010 / DX10
(20/50 Ω stator)
Comet
XSTAR
(25/50 Ω stator)
Dunlee (Philips)
DA10 series
(20/50 Ω stator)
240 VAC
240 VAC
1.5 sec
1.5 sec
1.5 sec
1.5 sec
2.0 sec
30 µF
30 µF
30 µF
30 µF
LSS PART NO.
(See note 2)
903836-00
903836-00
903836-00
903836-00
903836-00
40 µF
903836-01
(50 or 60 Hz) (prefer)
30 µF
903836-00
(60 Hz only)
CMP 150™
BOOST VOLTS
( AC SUPPLY)
120 VAC
BOOST
TIME
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
240 VAC
120 VAC
N/A
2.0 sec
N/A
240 VAC
2.0 sec
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
SHIFT CAP.
30 µF
LSS PART NO.
(See note 2)
908306-00
30 µF
908306-00
30 µF
908306-00
30 µF
908306-00
30 µF
908306-00
40 µF
908306-01
(50 or 60 Hz)
30 µF
(60 Hz only)
908306-00
Page 8
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
CMP 200® / CMP 200® DR
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
BOOST
VOLTS
BOOST
TIME
SHIFT CAP.
LSS PART NO.
(See note 2)
Dunlee (Philips)
DA90, DA90HS
(36/44 Ω stator)
(See note 4)
DU 1750
DU 3050
DU 2550
DU 33100
(See note 6)
DR1429 (DU304-0.6/1.2)
DR1436 (DU404-0.6/1.2)
DR1492 (DU604-0.6/1.2)
DR1494 (DU404-0.4/0.8)
240 VAC
1.5 sec
30 µF
903836-00
Dunlee (Philips)
DR1400 series
(20/50 Ω stator)
Dunlee (Philips)
DR1800 series
(19/22 Ω stator)
(See note 4)
DR1817
DR1833
Dunlee
(Picker / Philips)
PX1300 series
(3” anode)
“S” stator
(15/30 Ω
Dunlee
(Picker / Philips)
PX1400 series
(4” anode)
“S” stator
(15/30 Ω)
PX1302 (DU303-1.0/2.0)
PX1312 (DU303-0.6/1.2)
PX1351 (DU143-1.0/2.0)
PX1402 PX1412
PX1415 PX1425
PX1429 PX1431
PX1436 PX1456
PX1463 PX1472
PX1473 PX1475
PX1482 PX1483
PX1492 PX1494
DU404 (0.6/1.25)
240 VAC
240 VAC
240 VAC
240 VAC
2.0 sec
1.5 sec
2.0 sec
2.0 sec
CMP 150™
BOOST
BOOST
VOLTS
TIME
( AC SUPPLY)
120 VAC
N/A
240 VAC
1.5 sec
40 µF
903836-01
(50 or 60 Hz) (prefer)
120 VAC
N/A
240 VAC
2.0 sec
30µF
(60 Hz only)
903836-00
120 VAC
N/A
240 VAC
2.0 sec
30 µF
903836-00
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
2.0 sec
120 VAC
N/A
240 VAC
2.0 sec
30 µF
30 µF
903836-00
903836-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
SHIFT CAP.
LSS PART NO.
(See note 2)
30 µF
908306-00
40 µF
908306-01
(50 or 60 Hz)
30 µF
908306-00
(60 Hz only)
30 µF
908306-00
30 µF
908306-00
30 µF
908306-00
Page 9
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Dunlee
(Picker / Philips)
PX1400 series
(4” anode)
“S” stator
(15/30 Ω)
PX1402 (DU304-1.0/2.0)
PX1412 (DU304-0.6/1.2)
PX1415 (DU404-0.3/0.9)
PX1425
PX1429 (DU304-0.6/1.2)
PX1431 (DU304-0.7/1.4)
PX1436 (DU404-0.6/1.2)
PX1456 (DU404-0.3/1.0)
PX1463 (DU604-0.6/1.2)
PX1472 (DU604-0.3/1.0)
PX1473 (DU404-0.6/1.2)
PX1475 (DU754-0.6/1.2)
PX1482 (DU404-0.6/1.0)
PX1483 (DU404-0.3/1.0)
PX1492 (DU604-0.6/1.2)
PX1494 (DU404-0.4/0.8
240 VAC
MX-75
240 VAC
2.0 sec
30 µF
LSS PART NO.
(See note 2)
CMP 150™
BOOST VOLTS
( AC SUPPLY)
BOOST
TIME
SHIFT CAP.
LSS PART NO.
(See note 2)
903836-00
120 VAC
N/A
30 µF
908306-00
240 VAC
2.0 sec
120 VAC
N/A
40 µF
908306-01
240 VAC
1.5 sec
120 VAC
N/A
40 µF
908306-01
240 VAC
1.5 sec
Eureka (see Varian)
GE Maxiray 75
(3” anode)
(23/23 Ω equal
impedance
“E” stator)
* Dunlee tube
GE Maxiray 100
(4” anode)
(23/23 Ω equal
impedance “E”
stator)
1.5 sec
40 µF
903836-01
DU303 (1.0/2.0) *
DU303 (0.6/1.2) *
MX-100
240 VAC
1.5 sec
40 µF
903836-01
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 10
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Gilardoni
Rotagil S/AS
AR11-30
AR20-50
AR30-00-1
AR30-60
AR30-100
AR40-100
AR9000-1
AR9000-2
Hangzhou Kailong (see Kailong RADII)
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
240 VAC 1.5 sec 30 µF
Hangzhou Wandong WANRAY
(Polygala / Yuan Zhi)
XD51-20.40/125 (LQ16-XA2)
XD51-10.40/125 (LQ16-XA4)
XD52-30.50/125 (LQ17-XB1)
XD57-20.50/150 (LQ16-XD57)
IAE
RTM 70H
X20P
C30, C31
X22
(25/62 Ω stator)
X40
IAE
X20P
X22
C40
X40
(33/64 Ω stator)
240 VAC
IAE
C52, C52 Super
C100
C352
(20/40 Ω stator)
240 VAC
RTC 600HS
RTC 700HS
RTC 1000HS
RTM 78H / HS
RTM 90H / HS
RTM 92H / HS
RTM 101H / HS
RTM 102H / HS
RTM 780H / HS
RTM 782H / HS
X39
X40
X50H X50AH
240 VAC
240 VAC
1.5 sec
1.5 sec
1.5 sec
1.5 sec
LSS PART NO.
(See note 2)
903836-00
40 µF
903836-01
(50 Hz only)
30 µF
20 µF
903836-00
903836-02
CMP 150™
BOOST VOLTS
( AC SUPPLY)
120 VAC
BOOST
TIME
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
SHIFT CAP.
30 µF
LSS PART NO.
(See note 2)
908306-00
40 µF
(50 Hz only)
908306-01
30 µF
908306-00
20 µF
908306-02
30 µF
903836-00
(prefer)
120 VAC
N/A
30 µF
908306-00
(prefer)
40 µF
903836-01
240 VAC
1.5 sec
40 µF
908306-01
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 11
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
30 µF
240 VAC 1.5 sec
CMP 150™
BOOST VOLTS
( AC SUPPLY)
120 VAC
BOOST
TIME
N/A
240 VAC
1.5 sec
30 µF
903836-00
(50 or 60 Hz) (prefer)
120 VAC
N/A
240 VAC
1.5 sec
40 µF
(50 Hz only)
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
N/A
1.8 sec
N/A
1.8 sec
N/A
30 µF
908306-00
(50 or 60 Hz) (prefer)
40 µF
908306-01
(50 Hz only)
903836-00
120 VAC
240 VAC
120 VAC
240 VAC
120 VAC
30 µF
908306-00
903836-00
120 VAC
N/A
30 µF
908306-00
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Kailong RADII
H1064
H1065
H1066
(25/62 Ω stator)
KL64
KL65
KL66
Kailong RADII
H1074
H1075
H1080
H1083
H1086
(20/50 Ω stator)
KL74
KL75
KL80
KL80A
KL83
KL86
Kailong RADII
H1076
KL76
240 VAC 1.5 sec
20 µF
Kailong RADII
H2100
(20/50 Ω stator)
KL100
240 VAC 1.8 sec
30 µF
903836-00
(50 or 60 Hz) (prefer)
903836-01
40 µF
(50 Hz only)
Machlett (Varex)
Dynamax (DX) 52
“R” stator (16/50 Ω)
Machlett (Varex)
Dynamax (DX) 62
“R” stator
(23/56 Ω)
Dynamax (DX) 62U
configured as
“STD” or “R” stator
(15/36 Ω)
A-102
A-132
A-142
240 VAC 1.5 sec
30 µF
A-192B
A-196
240 VAC 1.5 sec
30 µF
A-197
A-256
A-272
A-282
240 VAC 1.5 sec
A-286
A-292
A-482
A-486
240 VAC 1.5 sec
30 µF
LSS PART NO.
(See note 2)
903836-00
903836-01
903836-02
30 µF
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
SHIFT CAP.
30 µF
LSS PART NO.
(See note 2)
908306-00
30 µF
908306-00
(50 or 60 Hz) (prefer)
40 µF
(50 Hz only)
908306-01
20 µF
908306-02
Page 12
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Philips
ROT350
ROT351
RO 1230
RO 1750 / DU 1750*
RO 3050 / DU 3050*
SRO 2050
SRO 2250
GD6 3050 **
PG-256 ***
PG-292 ***
RAD-34 ***
RAD-50 ***
(36/44 Ω stator)
(See note 4)
* Dunlee tube
(See note 6)
** NAGO tube
*** Varex (Varian)
tube
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
240 VAC 1.5 sec 30 µF
LSS PART NO.
(See note 2)
903836-00
CMP 150™
BOOST VOLTS
( AC SUPPLY)
120 VAC
BOOST
TIME
N/A
240 VAC
1.5 sec
120 VAC
(See note 10)
1.5 sec
240 VAC
N/A
120 VAC
(See note 10)
240 VAC
1.5 sec
SHIFT CAP.
30 µF
LSS PART NO.
(See note 2)
908306-00
30 µF
908306-00
30 µF
908306-00
30 µF
908306-00
Picker (see Dunlee)
Shimadzu CIRCLEX P18DE-85
RX80
RX81
RX82
(7.5/35 Ω stator)
120 VAC
Shimadzu CIRCLEX P38E
RX100
120 VAC
Shimadzu CIRCLEX P18DK
RX150
(8.4/24 Ω stator)
120 VAC
1.5 sec
1.5 sec
1.5 sec
30 µF
30 µF
30 µF
903836-00
903836-00
903836-00
120 VAC
(See note 10)
240 VAC
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
N/A
1.5 sec
N/A
Rev. AF
Page 13
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Siemens Opti-150
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian) tube
Siemens (SXVT)
RAY-6_1
RAY-8(S)_1
RAY-12(S)_1
(1-phase only)
(20/50 Ω stator)
20/40
30/52R
SG-796B *
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
120 VAC 2.0 sec 40 µF
SV 150/12/40-S
SV 150/25/48-S
SV 150/22/54-S
SV 150/33/78R-S
240 VAC
Siemens (SXVT)
RAY-13(S)_1
RAY-14(S)_1
(1-phase only)
(20/50 Ω stator)
SV 150/33/78R-S
240 VAC
Siemens (SXVT)
SDR
(1-phase only)
(20/50 Ω stator)
SDR 150/30/50-1
Siemens (SXVT)
SV 150
“S” stator
(1-phase only)
(14/18 Ω)
SV 150/40/80C-100_1
(See note 7)
240 VAC
120 VAC
1.5 sec
1.5 sec
1.5 sec
2.0 sec
30 µF
30 µF
30 µF
40 µF
LSS PART NO.
(See note 2)
903836-01
903836-00
903836-00
903836-00
903836-01
CMP 150™
BOOST VOLTS
( AC SUPPLY)
120 VAC
(See note 10)
BOOST
TIME
2.0 sec
240 VAC
N/A
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
(See note 10)
240 VAC
2.0 sec
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
SHIFT CAP.
40 µF
LSS PART NO.
(See note 2)
908306-01
30 µF
908306-00
30 µF
908306-00
30 µF
908306-00
40 µF
908306-01
N/A
Rev. AF
Page 14
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
CMP 150™
BOOST VOLTS
( AC SUPPLY)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba Rotanode
Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The
housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each
having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing
used. For ease of use, the Toshiba tubes are sorted by Tube Insert Type (“E” number) within this table. The installer must confirm the stator
type for the planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the
appropriate table entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same row within
this table.
20 µF
E7132
240 VAC 1.5 sec 20 µF
903836-02
120 VAC
N/A
908306-02
E7239
E7240
E7242
240 VAC
1.5 sec
(See note 8)
Toshiba
XH-121
XH-126
XH-150
XS-AV stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-106V
XH-180
XH-181
XS-AL stator
(9.4/28.3 Ω)
E7252
(See note 8)
120 VAC
1.5 sec
40 µF
903836-01
(50 or 60 Hz) (prefer)
30 µF
(60 Hz only)
Toshiba
XH-106V
XH-180
XH-181
XS-RA stator
(27.5/58 Ω)
E7252
(See note 8)
Toshiba
XH-157
XS-RB stator
(20.2/38 Ω)
E7254
E7255
(See note 8)
240 VAC
240 VAC
1.5 sec
1.5 sec
LSS PART NO.
(See note 2)
20 µF
30 µF
903836-00
903836-02
903836-00
BOOST
TIME
120 VAC
(See note 10)
240 VAC
1.5 sec
120 VAC
(See note 10)
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
SHIFT CAP.
LSS PART NO.
(See note 2)
40 µF
903836-01
(prefer)
30 µF
908306-00
20 µF
908306-02
30 µF
908306-00
N/A
N/A
Rev. AF
Page 15
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba
XH-121
XH-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-106V
XH-180
XS-AL stator
(9.4/28.3 Ω)
E7299
(See note 8)
Toshiba
XH-121
XS-BA stator
(18/47.5 Ω)
E7843
(See note 8)
Toshiba
XH-121
XH-126
XH-150
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-112V-2
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-121
XS-AV stator
(27.5/58 Ω)
E7861
(See note 8)
E7813
(See note 8)
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
20 µF
240 VAC 1.5 sec
120 VAC
120 VAC
240 VAC
1.5 sec
2.5 sec
1.5 sec
LSS PART NO.
(See note 2)
903836-02
CMP 150™
BOOST VOLTS
( AC SUPPLY)
120 VAC
BOOST
TIME
N/A
240 VAC
1.5 sec
1.5 sec
SHIFT CAP.
20 µF
LSS PART NO.
(See note 2)
908306-02
40 µF
(50 or 60 Hz)
903836-01
(prefer)
30 µF
(60 Hz only)
903836-00
40 µF
(50 or 60 Hz)
903836-01
(prefer)
120 VAC
(See note 10)
240 VAC
120 VAC
(See note 10)
240 VAC
120 VAC
(See note 10)
30 µF
(60 Hz only)
903836-00
240 VAC
N/A
30 µF
(60 Hz only)
908306-00
20 µF
903836-02
120 VAC
N/A
20 µF
908306-02
240 VAC
1.5 sec
N/A
1.5 sec
N/A
1.5 sec
40 µF
903836-01
(50 or 60 Hz) (prefer)
30 µF
(60 Hz only)
908306-00
903836-01
40 µF
(50 or 60 Hz) (prefer)
E7864
(See note 8)
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
E7865
(See note 8)
240 VAC
1.5 sec
20 µF
903836-02
120 VAC
N/A
20 µF
908306-02
240 VAC
1.5 sec
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 16
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba
XH-112V-2
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-121
XS-AV-stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-121
XH-126
XH-150
XS-AL stator
(9.4/28.3 Ω)
E7869
(See note 8)
Toshiba
XH-121
XH-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
E7886
(See note 8)
E7876
(See note 8)
E7884
(See note 8)
CMP 200® / CMP 200® DR
BOOST
BOOST
SHIFT CAP.
VOLTS
TIME
N/A
N/A
N/A
240 VAC
120 VAC
1.5 sec
1.5 sec
20 µF
1.5 sec
CMP 150™
BOOST VOLTS
( AC SUPPLY)
N/A
BOOST
TIME
N/A
SHIFT CAP.
N/A
LSS PART NO.
(See note 2)
N/A
903836-02
120 VAC
N/A
20 µF
908306-02
240 VAC
1.5 sec
120 VAC
(See note 10)
240 VAC
120 VAC
(See note 10)
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
903836-01
40 µF
(50 or 60 Hz) (prefer)
30 µF
(60 Hz only)
240 VAC
LSS PART NO.
(See note 2)
N/A
20 µF
903836-00
903836-02
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
N/A
1.5 sec
40 µF
908306-01
(50 or 60 Hz) (prefer)
30 µF
(60 Hz only)
908306-00
20 µF
908306-02
N/A
Rev. AF
Page 17
Table 1A: X-ray Tube Types of Low-Speed Starter for CMP 200® / CMP 150™ Generators
CPI Canada Inc
TABLE 1A: TUBE TYPES – LOW SPEED STARTER – CMP 200® / CMP 150™ GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Varex (Varian)
B-100
“STD” stator
(16/50 Ω)
Varex (Varian)
B-130
B-130H
B-135H
B-146
B-147
B-150
B-150H
Std “R” stator
(16/50 Ω)
Varex
(Varian / Eureka )
Diamond
Emerald
LEO
Std “R” stator
(20/50 Ω)
Varex
(Varian / Eureka)
Sapphire
Std “R” stator
(20/50 Ω)
CMP 200® / CMP 200® DR
CMP 150™
BOOST
VOLTS
BOOST
TIME
SHIFT CAP.
LSS PART NO.
(See note 2)
BOOST VOLTS
( AC SUPPLY)
BOOST
TIME
SHIFT CAP.
LSS PART NO.
(See note 2)
A-102
A-132 / A-134*
A-142 / A-144*
A-145
* (See note 9)
A-152
A-182 / A-184*
A-192 / A-194*
A-195
A-196 A-197
A-272 / A-274*
A-277 / A-278*
A-282 / A-284*
A-286
A-292 / A-294*
A-482
G-242
G-256
G-292 / G-294*
G-296
G-892
G-896
* (See note 9)
RAD-8
RAD-12
RAD-13
RAD-14
RAD-68
240 VAC
1.5 sec
30 µF
903836-00
120 VAC
N/A
30 µF
908306-00
240 VAC
1.5 sec
120 VAC
N/A
30 µF
908306-00
240 VAC
1.5 sec
120 VAC
N/A
240 VAC
1.5 sec
120 VAC
N/A
30 µF
908306-00
240 VAC
1.5 sec
RAD-21
RAD-40
RAD-44
RAD-56
RAD-60
RAD-92
RAD-94
240 VAC
120 VAC
N/A
30 µF
908306-00
240 VAC
1.5 sec
240 VAC
240 VAC
240 VAC
1.5 sec
30 µF
903836-00
1.5 sec
1.5 sec
1.5 sec
30 µF
30 µF
903836-00
903836-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 18
Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators
CPI Canada Inc
Low-Speed Starter Tube Select Table – Mobile DR Generators
Low-Speed-Starter X-ray Tube Types
Table 1B: Mobile DR Generators
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 19
Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators
CPI Canada Inc
TABLE 1B: TUBE TYPES – LOW SPEED STARTER – MOBILE DR GENERATORS
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
BOOST
VOLTS
BOOST
TIME
SHIFT CAP.
LSS PART NO.
(See note 2)
IAE
C30, C31
(25/62 Ω stator)
IAE
C40
(33/64 Ω stator)
X20P
X22
X40
240 VAC
1.5 sec
30 µF
904832-00
X20P
X22
X40
240 VAC
1.5 sec
30 µF
904832-00
Kailong RADII
H1064, H1065, H1066
(25/62 Ω stator)
KL64
KL65
KL66
240 VAC
1.5 sec
30 µF
904832-00
Kailong RADII
H1074, H1075
H1080, H1083, H1086
(20/50 Ω stator)
KL74, KL75
KL80, KL80A
KL83, KL86
240 VAC
1.5 sec
30 µF
904832-00
Kailong RADII
H1076
KL76
240 VAC
1.5 sec
30 µF
904832-00
Siemens (SXVT)
RAY-6_1
RAY-8(S)_1
RAY-12(S)_1
(1-phase only)
(20/50 Ω stator)
SV 150/12/40-S
SV 150/25/48-S
SV 150/22/54-S
SV 150/33/78R-S
240 VAC
1.5 sec
30 µF
904832-00
Siemens (SXVT)
RAY-13(S)_1
RAY-14(S)_1
(1-phase only)
(20/50 Ω stator)
SV 150/12/40-S
SV 150/25/48-S
SV 150/22/54-S
SV 150/33/78R-S
240 VAC
1.5 sec
30 µF
904832-00
Siemens (SXVT)
SDR
(1-phase only)
(20/50 Ω stator)
SDR 150/30/50-1
240 VAC
1.5 sec
30 µF
904832-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 20
Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators
CPI Canada Inc
TABLE 1B: TUBE TYPES – LOW SPEED STARTER – MOBILE DR GENERATORS (Continued)
TUBE TYPE
(HOUSING)
Toshiba Rotanode
Toshiba
XH-106V, XH-180, XH-181
XS-RA stator
(27.5/58 Ω)
TUBE TYPE
BOOST
BOOST
LSS PART NO.
SHIFT CAP.
(INSERT)
VOLTS
TIME
(See note 2)
Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the
complete assembly. The housing number is usually not shown on the identification label. A specific “E” number may be
available with several different stator types, each having different starter requirements. Furthermore, a specific stator type
may require different starter requirements depending on the housing used. For ease of use, the Toshiba tubes are sorted
by Tube Insert Type (“E” number) within this table. The installer must confirm the stator type for the planned tube(s). If the
stator type is not identified within the Toshiba documentation, match the stator winding resistance to the appropriate table
entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same row
within this table.
904832-00
30 µF
E7252
240 VAC
1.5 sec
(See note 1)
Toshiba
XH-121, XH-126, XH-150
XS-RA stator
(27.5/58 Ω)
E7861
(See note 1)
240 VAC
1.5 sec
30 µF
904832-00
Toshiba
XH-121
XS-AV-stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-121
XS-AV-stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-121, XH-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
E7865
(See note 1)
240 VAC
1.5 sec
30 µF
904832-00
E7876
(See note 1)
240 VAC
1.5 sec
30 µF
904832-00
E7886
(See note 1)
240 VAC
1.5 sec
30 µF
904832-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 21
Table 1B: X-ray Tube Types of Low-Speed Starter for Mobile DR Generators
CPI Canada Inc
TABLE 1B: TUBE TYPES – LOW SPEED STARTER – MOBILE DR GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
BOOST
VOLTS
BOOST
TIME
SHIFT CAP.
LSS PART NO.
(See note 2)
Varex (Varian)
Diamond
Emerald
LEO
Std “R” stator
(20/50 Ω)
RAD-8
RAD-12
RAD-13
RAD-14
RAD-68
RAD-74
240 VAC
1.5 sec
30 µF
904832-00
NOTE 1: Complete Toshiba x-ray tube assemblies include the suffix “X, “FX”, “GX”,
or “JX”, which are interchangeable with respect to rotor and anode characteristics.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 22
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
Low-Speed Starter Tube – Indico IQ® / Indico 100® Generators
Low-Speed-Starter X-ray Tube Types
Table 1C: Indico IQ® / Indico 100® Generators
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 23
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100®® GENERATORS
INDICO IQ®
CODE,
OUTPUT SHIFT
Switches
FREQ.
CAP.
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CGR (GE)
Statorix
200A/240/260
(50/110 Ω stator)
MN640
M641
MN641
M643
Type 28
(00111)
50 Hz
CGR (GE)
Statorix 550
(50/110 Ω stator)
RN620
RN621
Type 28
(00111)
Chirana
Rotax KA 125
(20/50 Ω stator)
RIK-T 0.8/2.0 12/50
RIK-T 1.2/2.0 30/50
Chirana
Rotax KA 125
(20/20/20 Ω stator)
INDICO 100®
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
12.5µF
905135-41
240 VAC
2.5 sec
94 VAC
12.5µF
732752-01
50 Hz
12.5µF
905135-41
240 VAC
2.5 sec
94 VAC
12.5µF
732752-01
Type 21
(10101)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
73 VAC
30 µF
732752-00
N/A
N/A
N/A
N/A
240 VAC
1.5 sec
73 VAC
100µF,
52 mH
732752-04
(1........5)
Comet
DO7 / DX7
(25/50 Ω stator)
DX7
DX71HS
(See note 3)
Type 0
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Comet
DO9 / DX9
(20/50 Ω stator)
DX9
DX91H / HS
DX92H / HS
DX93H / HS
DX94HS
DX96HS
DX97HS
(See note 3)
Type 0
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 24
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Comet
D010 / DX10
(20/50 Ω stator)
DX10H / HS
DX101H / HS
DX104HS
DX105HS
DX106HS
(See note 3)
DX700HS
(See note 3)
Type 0
(00000)
Comet
XSTAR
(25/50 Ω stator)
XSTAR8 (XST-8)
XSTAR74 (XST-74)
XSTAR-WHIS
Dunlee (Philips)
DA10 series
(20/50 Ω stator)
Comet
DO700 / DX700
(25/50 Ω stator)
Dunlee (Philips)
DA90, DA90HS
(36/44 Ω stator)
(See note 4)
Dunlee (Philips)
DA90, DA90HS
(9/11 Ω stator)
(See note 5)
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Type 1
(10000)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
Type 0
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
DA1002 (DU304-1.0/2.0) Type 4
DA1029 (DU304-0.6/1.2) (00100)
DA1036 (DU404-0.6/1.2)
DA1083 (DU404-0.3/1.0)
DA1092 (DU604-0.6/1.2)
DA1094 (DU404-0.4/0.8)
DU404 (0.6/1.25)
DU 1750
Type 9
DU 3050
(10010)
DU 2550
DU 33100
(See note 6)
DU 1750
Type 9
DU 3050
(10010)
DU 2550
DU 33100
(See note 6)
60 Hz
33 µF
30 µF
40 µF
905135-41
905135-43
905135-44
240 VAC
2.5 sec
52 VAC
45 µF
732752-02
(50 or 60 Hz) (prefer)
Switches
(1........5)
FREQ.
30 µF
(60 Hz only)
732752-00
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 25
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Dunlee
(Philips)
DR1400 series
(20/50 Ω stator)
DR1429 (DU304-0.6/1.2) Type 4
DR1436 (DU404-0.6/1.2) (00100)
DR1492 (DU604-0.6/1.2)
DR1494 (DU404-0.4/0.8)
Switches
(1........5)
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
60 Hz
33 µF
30 µF
40 µF
905135-41
905135-43
905135-44
240 VAC
2.5 sec
52 VAC
45 µF
732752-02
(50 or 60 Hz) (prefer)
30 µF
732752-00
(60 Hz only)
FREQ.
LSS PART NO.
(See note 2)
Dunlee (Philips) DR1817
DR1800 series DR1833
(19/22 Ω stator)
(See note 4)
Type 9
(10010)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Dunlee (Philips) DR1817
DR1800 series DR1833
(9/11 Ω stator)
(See note 5)
Type 9
(10010)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
PX1302 (DU303-1.0/2.0) Type 7
PX1312 (DU303-0.6/1.2) (11100)
PX1351 (DU143-1.0/2.0)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
2.5 sec
52 VAC
30 µF
732752-00
PX1402 (DU304-1.0/2.0) Type 1
PX1412 (DU304-0.6/1.2) (10000)
PX1415 (DU404-0.3/0.9)
PX1429 (DU304-0.6/1.2)
PX1431 (DU304-0.7/1.4)
PX1436 (DU404-0.6/1.2)
PX1456 (DU404-0.3/1.0)
PX1463 (DU604-0.6/1.2)
PX1472 (DU604-0.3/1.0)
PX1473 (DU404-0.6/1.2)
PX1475 (DU754-0.6/1.2)
PX1482 (DU404-0.6/1.0)
PX1483 (DU404-0.3/1.0)
PX1492 (DU604-0.6/1.2)
PX1494 (DU404-0.4/0.8)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
2.5 sec
52 VAC
30 µF
732752-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Dunlee
(Picker /
Philips)
PX1300 series
(3” anode)
(“S” stator,
15/30 Ω)
Dunlee
(Picker /
Philips)
PX1400 series
(4” anode)
(“S” stator,
15/30 Ω)
Page 26
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
INDICO IQ®
CODE,
OUTPUT
Switches
(1........5)
FREQ.
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
PX1402 (DU304-1.0/2.0) Type 1
PX1412 (DU304-0.6/1.2) (10000)
PX1415 (DU404-0.3/0.9)
PX1425
PX1429 (DU304-0.6/1.2)
PX1431 (DU304-0.7/1.4)
PX1436 (DU404-0.6/1.2)
PX1456 (DU404-0.3/1.0)
PX1463 (DU604-0.6/1.2)
PX1472 (DU604-0.3/1.0)
PX1473 (DU404-0.6/1.2)
PX1475 (DU754-0.6/1.2)
PX1482 (DU404-0.6/1.0)
PX1483 (DU404-0.3/1.0)
PX1492 (DU604-0.6/1.2)
PX1494 (DU404-0.4/0.8)
Eureka (see Varex/Varian)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
2.5 sec
52 VAC
30 µF
732752-00
GE Maxiray 75
(3” anode)
(23/23 Ω equal
impedance “E”
stator)
Type 14
(01110)
60 Hz
45 µF
40 µF
905135-41
905135-44
240 VAC
1.5 sec
73 VAC
45 µF
732752-02
Type 20
(00101)
60 Hz
45 µF
40 µF
905135-41
905135-44
240 VAC
1.5 sec
73 VAC
45 µF
732752-02
Dunlee
(Picker / Philips)
PX1400 series
(4” anode)
“S” stator (15/30
Ω)
* Dunlee tube
GE Maxiray 100
(4” anode)
(23/23 Ω equal
impedance “E”
stator)
MX-75
DU303 (1.0/2.0) *
DU303 (0.6/1.2) *
MX-100
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 27
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
INDICO IQ®
CODE,
OUTPUT
Switches
(1........5)
FREQ.
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
Gilardoni
Rotagil S/AS
AR11-30
AR20-50
AR30-00-1
AR30-60
AR30-100
AR40-100
AR9000-001
AR9000-002
Hangzhou Kailong (see Kailong RADII)
Type 22
(01101)
50 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Hangzhou Wandong WANRAY
(Polygala / Yuan Zhi)
XD51-20.40/125 (LQ16-XA2)
XD51-10.40/125 (LQ16-XA4)
XD52-30.50/125 (LQ17-XB1)
XD57-20.50/150 (LQ16-XD57)
IAE
RTM 70H
X20P
C30, C31
X22
(25/62 Ω stator)
X40
IAE
X39
X40
C40
X42
(33/64 Ω stator)
Type 0
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
45 µF
(50 Hz only)
732752-02
Type 10
(01010)
50 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Type 27
(11011)
50 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
(prefer)
45 µF
732752-02
IAE
C52, C52 Super
C100
C352
(20/40 Ω stator)
RTC 600HS
RTC 700HS
RTC 1000HS
RTM 78H / HS
RTM 90H / HS
RTM 92H / HS
RTM 101H / HS
RTM 102H / HS
RTM 780H / HS
RTM 782H / HS
X39
X40
X50H X50AH
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 28
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
INDICO 100®
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Kailong RADII
H1064
H1065
H1066
(25/62 Ω stator)
KL64
KL65
KL66
Type 0
(00000)
Kailong RADII
H1074
H1075
H1080
H1083
H1086
(20/50 Ω stator)
KL74
KL75
KL80
KL80A
KL83
KL86
Type 0
(00000)
Kailong RADII
H1076
KL76
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
240 VAC
1.5 sec
Kailong RADII
H2100
(20/50 Ω stator)
KL100
Type 21
(10101)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
2.5 sec
Machlett (Varex)
Dynamax (DX) 52
“R” stator (16/50 Ω)
Machlett (Varex)
Dynamax (DX) 62
“R” stator
(23/56 Ω)
Dynamax (DX) 62U
configured as
“STD” or “R” stator
(15/36 Ω)
A-102
A-132
A-142
Type 0
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
A-192B
A-196
A-197
A-256
A-272
A-282
A-286
A-292
A-482
A-486
Type 0
(00000)
Type 1
(10000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
Switches
(1........5)
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
(50 or 60 Hz)
732752-00
(prefer)
45 µF
(50 Hz only)
732752-02
52 VAC
30 µF
732752-00
73 VAC
30 µF
(50 or 60 Hz)
732752-00
(prefer)
732752-02
52 VAC
45 µF
(50 Hz only)
30 µF
52 VAC
30 µF
732752-00
FREQ.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
732752-00
Page 29
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Philips
ROT350
ROT351
RO 1230
RO 1750 / DU
1750*
RO 3050 / DU
3050*
SRO 2050
SRO 2250
GD6 3050 **
PG-256 ***
PG-292 ***
RAD-34 ***
RAD-50 ***
RO 1230
RO 1750 / DU
1750*
RO 3050 / DU
3050*
SRO 2050
SRO 2250
GD6 3050 **
PG-256 ***
PG-292 ***
RAD-34 ***
RAD-50 ***
SRM 0310
(36/44 Ω stator)
(See note 4)
* Dunlee tube
(See note 6)
** NAGO tube
*** Varex (Varian)
tube
Philips
ROT350
ROT351
(9/11 Ω stator)
(See note 5)
* Dunlee tube
(See note 6)
** NAGO tube
*** Varex (Varian)
tube
Philips
ROT500
ROT501
(9/11 Ω stator)
INDICO IQ®
CODE,
OUTPUT
Switches
(1........5)
FREQ.
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
Type 9
(10010)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Type 9
(10010)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
Type 30
(01111)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
Picker (see Dunlee)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 30
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Shimadzu CIRCLEX
RX80
RX81
RX82
(7.5/35 Ω stator)
P18DE-85
Type 16
(00001)
Shimadzu CIRCLEX
RX100
P38E
Shimadzu CIRCLEX
RX150
(8.4/24 Ω stator)
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
RUN
VOLTS
SHIFT CAP.
LSS PART NO.
(See note 2)
60 Hz
33 µF
30 µF
905135-41
905135-43
120 VAC
1.5 sec
(See note 10)
52 VAC
30 µF
732752-00
Type 16
(00001)
60 Hz
33 µF
30 µF
905135-41
905135-43
120 VAC
1.5 sec
(See note 10)
52 VAC
30 µF
732752-00
P18DK
Type 16
(00001)
60 Hz
33 µF
30 µF
905135-41
905135-43
120 VAC
1.5 sec
(See note 10)
52 VAC
30 µF
732752-00
Siemens Opti-150
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian) tube
Siemens SV 150
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian) tube
Siemens (SXVT)
RAY-6_1
RAY-8(S)_1
RAY-12(S)_1
(1-phase only)
(20/50 Ω stator)
20/40
30/52R
SG-796B *
N/A
N/A
N/A
N/A
120 VAC
2.5 sec
(See note 10)
52 VAC
45 µF
732752-02
SV 150/40/80C-100L Type 23
SG-256B *
(11101)
50 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
SV 150/12/40-S
SV 150/25/48-S
SV 150/22/54-S
SV 150/33/78R-S
Type 31
(11111)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
73 VAC
30 µF
732752-00
Siemens (SXVT)
RAY-13(S)_1
RAY-14(S)_1
(1-phase only)
(20/50 Ω stator)
SV 150/33/78R-S
Type 31
(11111)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
73 VAC
30 µF
732752-00
Switches
(1........5)
FREQ.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
BOOST
TIME
Rev. AF
Page 31
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Siemens (SXVT)
SDR
(1-phase only)
(20/50 Ω stator)
SDR 150/30/50-1
Siemens (SXVT)
SV 150
“S” stator
(1-phase only)
(14/18 Ω)
Toshiba Rotanode
SV 150/40/80C-100_1
Toshiba
XH-112V-2
XH-112W
XS-AA stator
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-112V
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-121
XH-126
XH-150
XS-AV stator
XS-RA stator
(27.5/58 Ω)
(See note 7)
INDICO IQ®
CODE,
OUTPUT
Switches
FREQ.
(1........5)
Type 31
60 Hz
(11111)
Type 23
(11101)
50 Hz
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT CAP.
LSS PART NO.
(See note 2)
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
73 VAC
30 µF
732752-00
33 µF
30 µF
905135-41
905135-43
120 VAC
2.5 sec
(See note 10)
52 VAC
45 µF
732752-02
Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The
housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each
having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing
used. For ease of use, the Toshiba tubes are sorted by Tube Insert Type (“E” number) within this table. The installer must confirm the stator
type for the planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the
appropriate table entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same row within
this table.
N/A
N/A
33 µF
DRX-3624H
Type 17
60 Hz
905135-41
N/A
N/A
N/A
DRX-3724H
30 µF
(10001)
905135-43
DRX-4634H
E7100
Type 17
(10001)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
E7132
E7239
E7240
E7242
(See note 8)
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 32
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
INDICO 100®
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba
XH-112V-2
XS-AA stator
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-106V
XH-180
XH-181
XS-AL stator
(9.4/28.3 Ω)
E7250
(See note 8)
Type 17
(10001)
E7252
(See note 8)
Type 29
(10111)
Toshiba
XH-106V
XH-180
XH-181
XS-RA stator
(27.5/58 Ω)
E7252
(See note 8)
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
240 VAC
1.5 sec
Toshiba
XH-157
XS-RB stator
(20.2/38 Ω)
E7254
E7255
(See note 8)
Type 21
(10101)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
Toshiba
XH-121
XH-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-106V
XH-180
XS-AL stator
(9.4/28.3 Ω)
E7299
(See note 8)
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
240 VAC
E7813
(See note 8)
Type 29
(10111)
60 Hz
45 µF
40 µF
905135-41
905135-44
Switches
(1........5)
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
60 Hz
45 µF
40 µF
905135-41
905135-44
120 VAC
1.5 sec
(See note 10)
52 VAC
45 µF
(50 or 60 Hz)
732752-02
(prefer)
30 µF
(60 Hz only)
732752-00
94 VAC
30 µF
732752-00
1.5 sec
52 VAC
30 µF
(60 Hz only)
732752-00
1.5 sec
52 VAC
30 µF
732752-00
1.5 sec
120 VAC
(See note 10)
52 VAC
45 µF
(50 or 60 Hz)
732752-02
(prefer)
30 µF
(60 Hz only)
732752-00
FREQ.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 33
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
INDICO 100®
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba
XH-121
XS-BA stator
(18/47.5 Ω)
E7843
(See note 8)
Toshiba
XH-121
XH-126
XH-150
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-112V-2
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-121
XS-AV stator
(27.5/58 Ω)
Toshiba
XH-112V-2
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-121
XS-AV-stator
XS-RA stator
(27.5/58 Ω)
E7861
(See note 8)
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
240 VAC
1.5 sec
E7864
(See note 8)
Type 17
(10001)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
E7865
(See note 8)
Type 19
(11001)
60 Hz
25 µF
20 µF
905135-41
905135-42
E7869
(See note 8)
Type 17
(10001)
60 Hz
33 µF
30 µF
E7876
(See note 8)
Type 19
(11001)
60 Hz
25 µF
20 µF
Switches
(1........5)
Type 0
(00000)
FREQ.
60 Hz
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
33 µF
30 µF
905135-41
905135-43
120 VAC
2.5 sec
(See note 10)
52 VAC
45 µF
(50 or 60 Hz)
732752-02
(prefer)
30 µF
(60 Hz only)
732752-00
94 VAC
30 µF
732752-00
N/A
N/A
N/A
N/A
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
905135-41
905135-42
240 VAC
1.5 sec
94 VAC
30 µF
732752-00
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
BOOST
TIME
Rev. AF
Page 34
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba
XH-121
XH-126
XH-150
XS-AL stator
(9.4/28.3 Ω)
E7884
(See note 8)
Toshiba
XH-121
XH-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
E7886
(See note 8)
INDICO IQ®
CODE,
OUTPUT
Switches
(1........5)
Type 29
(10111)
Type 19
(11001)
FREQ.
60 Hz
60 Hz
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
45 µF
40 µF
905135-41
905135-44
1.5 sec
120 VAC
(See note 10)
25 µF
20 µF
905135-41
905135-42
240 VAC
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
BOOST
TIME
1.5 sec
Rev. AF
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
52 VAC
45 µF
(50 or 60 Hz)
732752-02
(prefer)
30 µF
(60 Hz only)
732752-00
30 µF
732752-00
94 VAC
Page 35
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Varex (Varian)
B-100
“STD” stator
(16/50 Ω)
A-102
A-132 / A-134*
A-142 / A-144*
A-145
* (See note 9)
A-152
A-182 / A-184*
A-192 / A-194*
A-195
A-196 A-197
* (See note 9)
A-272 / A-274*
A-277 / A-278*
A-282 / A-284*
A-286
A-292 / A-294*
A-482
G-242 G-256
G-292 / G-294*
G-296
G-892 G-896
* (See note 9)
G-1077
G-1078
G-1080
G-1082 / G-1084*
G-1086 / G-1087*
G-1092 / G-1094*
* (See note 9)
Varex (Varian)
B-130
B-130H
B-135H
B-146
B-147
B-150
B-150H
Std “R” stator
(16/50 Ω)
Varex (Varian)
B-160
B-160H
B-165H
“R” Stator
(16/50 Ω)
INDICO IQ®
CODE,
OUTPUT
Switches
(1........5)
FREQ.
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
Type 2
(01000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Type 2
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
Type 1
(10000)
Type 15
(11110)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 36
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
TABLE 1C: TUBE TYPES – LOW SPEED STARTER – INDICO IQ® /INDICO 100® GENERATORS (Continued)
INDICO IQ®
CODE,
OUTPUT
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Varex (Varian)
B-180
B-180H
B-185
B-185H
“R” Stator
(16/50 Ω)
Varex (Varian)
B-199
“S” Stator
(15/18 Ω)
Varex
(Varian/ Eureka)
Diamond
Emerald
LEO
Std “R” stator
(20/50 Ω)
Varex
(Varian / Eureka)
Sapphire
Std “R” stator
(20/50 Ω)
G-1582
G-1592
G-1593
Type 15
(11110)
SG-1096
INDICO 100®
SHIFT
CAP.
LSS PART NO.
(See notes 1,2)
BOOST
VOLTS
BOOST
TIME
RUN
VOLTS
SHIFT
CAP.
LSS PART NO.
(See note 2)
60 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
Type 18
(01001)
50 Hz
33 µF
30 µF
905135-41
905135-43
N/A
N/A
N/A
N/A
N/A
RAD-8
RAD-12
RAD-13
RAD-14
RAD-68
RAD-74
Type 0
(00000)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
RAD-21
RAD-40
RAD-44
RAD-56
RAD-60
RAD-92
RAD-94
Type 5
(10100)
60 Hz
33 µF
30 µF
905135-41
905135-43
240 VAC
1.5 sec
52 VAC
30 µF
732752-00
Switches
(1........5)
FREQ.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 37
Table 1C: X-ray Tube Types of Low-Speed Starter for Indico IQ® / Indio 100® Generators
CPI Canada Inc
Notes for Tables 1A, 1B, and 1C
NOTE 2: Where more than one Low Speed Starter part number is referenced for a
particular tube type and generator combination, any of these is acceptable.
In some cases, a preferred part number is identified.
NOTE 3: Comet tube inserts with the prefix “DI” and “DX” are interchangeable.
NOTE 4: Philips / Dunlee housing with the windings connected in series (low-speed
/ high impedance configuration).
NOTE 5: Philips / Dunlee housing with the windings connected in parallel (highspeed / low impedance configuration).
NOTE 6: Dunlee tube inserts with the prefix “DU” and Philips tube inserts with the
prefix “RO” are interchangeable. Select the corresponding Philips insert type within
the Generator software if the Dunlee type is not available.
NOTE 7: Siemens (SXVT) housing with the “phase-shifted” windings connected in
parallel (low impedance, high speed configuration).
NOTE 8: Complete Toshiba x-ray tube assemblies include the suffix “X, “FX”, “GX”,
or “JX”, which are interchangeable with respect to rotor and anode characteristics.
NOTE 9: These X-ray tubes incorporate a control grid. Grid control is currently not
supported by standard CPI generators. Connect the grid connection to Cathode
Common when using these tubes, and select the insert type within the Generator
software corresponding to the equivalent non-grid tube.
NOTE 10: The starter boost voltage required is 120 VAC. This is not available with
CMP 150™ High Frequency X-ray generators operating from a nominal 200-240V
AC line.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 38
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
Dual-Speed Starter Tube Select Table
Dual-Speed-Starter X-ray Tube Types
Table 2: Indico IQ® / Indico 100® / CMP 200®
Generators
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 39
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
The Dual Speed Starter (DSS) synthesizes its output frequencies independently of
the line frequency will operate all tubes at 60/180 Hz, unless a particular tube only
has published ratings to operate at 150 Hz. In this case, the starter will output
50/150 Hz when set for these tubes.
Unless indicated otherwise within Table 2, setting the switches to the code
indicated for the applicable Tube Type selects all required operating parameters.
The Dual Speed Starter part number corresponds to the AC line voltage (except
Indico IQ) and the configuration of phase-shift capacitors installed.
Reference notes are provided at the end of Table 2. Tube operating parameters
for each switch code are contained in Table 3 for reference.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 40
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPU
T FREQ.
H.S.
SHIFT
CAP.
L.S. SHIFT
CAP.
INDICO 100®
DSS PART NO.
(See note 2)
INDICO IQ®
DSS PART NO.
(See note 2)
CMP 200®
DSS PART NO.
(See note 2)
CGR (GE)
Statorix
200A/240/260
(50/110 Ω stator)
MN640
M641
MN641
M643
Type 28
(00111)
LOW SPEED
OPERATION ONLY
(See note 1)
50 Hz
N/A
12.5µF to
15.5µF
733317-15
735925-15
906894-15
(See note 10)
904925-23
901297-15
901298-15
(See note 10)
906672-23
MS740
MSN740
MSN742
RSN742
Type 28
(00111)
150 Hz
3 µF
12.5µF to
15.5µF
733317-15
735925-15
906894-15
(See note 10)
904925-23
901297-15
901298-15
(See note 10)
906672-23
RN620
RN621
Type 28
(00111)
50 Hz
N/A
12.5µF to
15.5µF
733317-15
735925-15
906894-15
(See note 10)
904925-23
901297-15
901298-15
(See note 10)
906672-23
150 Hz
3 µF
12.5µF to
15.5µF
CGR (GE)
Statorix 550
(50/110 Ω stator)
RSN722
Switches
(1........5)
Type 28
(00111)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Dual Speed Starters “-15” must
be jumper-configured to provide
the required 15.5 µF low speed
shift capacitance. Please see
note 10 before attempting to
use this Dual Speed Starter
configuration.
HIGH SPEED
OPERATION ONLY
(See note 1)
For Dual Speed operation,
contact factory.
LOW SPEED
OPERATION ONLY
(See note 1)
Dual Speed Starters “-15” must
be jumper-configured to provide
the required 15.5 µF low speed
shift capacitance. Please see
note 10 before attempting to
use this Dual Speed Starter
configuration.
HIGH SPEED
OPERATION ONLY
(See note 1)
For Dual Speed operation,
contact factory.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 41
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
Chirana
Rotax KA 125
(20/50 Ω stator,
single phase)
RIK-T
Type 21
(10101)
LOW SPEED
OPERATION ONLY
(See note 1)
Comet
DO7 / DX7
(25/50 Ω stator)
DX7
DX71HS
(See note 3)
Type 0
(00000)
Comet
DO9 / DX9
(20/50 Ω stator)
DX9
DX91H / HS
DX92H / HS
DX93H / HS
DX94HS
DX96HS
DX97HS
(See note 3)
DX10H
(See note 3)
Comet
DO10 / DX10
(20/50 Ω stator)
DX10HS
DX104HS
(See note 3)
DX105HS
DX106HS
(See note 3)
60 Hz
H.S.
SHIFT
CAP.
N/A
L.S.
SHIFT
CAP.
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 0
(00000)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 0
(00000)
LOW SPEED
OPERATION ONLY
(See note 1)
None
60 Hz
N/A
30 µF to
36 µF
50/150 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
None
50/150 Hz
6 µF
30 µF to
36 µF
Switches
(1........5)
Type 24
(00011)
Type 24
(10011)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Page 42
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
Comet
DO700 / DX700
(25/50 Ω stator)
DX700HS
(See note 3)
Dunlee (Philips)
DA10 series
(20/50 Ω stator)
Dunlee (Philips)
DA90 DA90HS
(36/44 Ω stator)
(See note 4)
Dunlee (Philips)
DA90
DA90HS
(9/11 Ω stator)
(See note 5)
Dunlee (Philips)
DR1400
(20/50 Ω stator)
OPERATING NOTES
OUTPUT
FREQ.
50/150 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 24
(10011)
None
DA1002
DA1029
DA1036
DA1083
DA1092
DA1094
DU404
(0.6/1.25)
DU 1750
DU 3050
DU 2550
DU 33100
(See note 6)
Type 4
(00100)
None
60/180 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 9
(10010)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30µF
DU 1750
DU 3050
(See note 6)
DU 2550
(See note 6)
DU 33100
(See note 6)
Type 9
(10010)
LOW SPEED
OPERATION ONLY
(See note 1)
None
60 Hz
N/A
37.5µF
60/180 Hz
12.5µF
37.5µF
180 Hz
12.5µF
37.5µF
DR1429
DR1436
DR1492
DR1494
DU404
(0.6/1.25)
Type 4
(00100)
HIGH SPEED
OPERATION ONLY
(See note 1)
None
60/180 Hz
6 µF
30 µF to
36 µF
Switches
(1........5)
Type 9
(10010)
Type 9
(10010)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-23
901297-13
901298-13
906672-21
906672-23
733317-16
735925-16
906894-16
733317-17
735925-17
906894-17
904925-21
904925-23
901297-16
901298-16
906672-21
906672-23
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Rev. AF
Page 43
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
OUTPUT
FREQ.
Dunlee (Philips)
DR 1800
(19/22 Ω stator)
(See note 4)
DR1817
Type 9
(10010)
LOW SPEED OPERATION
ONLY (See note 1)
Dunlee (Philips)
DR 1800
(9/11 Ω stator)
(See note 5)
DR1817
Type 9
(10010)
Type 9
(10010)
Type 9
(10010)
Type 7
(11100)
DR1825
DR1833
Dunlee
(Picker / Philips)
PX1300
3” anode
“S“ stator
(15/30 Ω)
PX1302
PX1312
PX1351
DU303 (1.0/2.0)
Dunlee
(Picker / Philips)
PX1400 series
4” anode
“S” stator
(15/30 Ω)
PX1402 PX1412
PX1415 PX1425
PX1429 PX1431
PX1436 PX1456
PX1463 PX1472
PX1473 PX1475
PX1482 PX1483
PX1492 PX1494
DU404 (0.6/1.25)
PX1402 PX1412
PX1415 PX1425
PX1429 PX1431
PX1436 PX1456
PX1463 PX1472
PX1473
PX1482 PX1483
PX1492 PX1494
DU404 (0.6/1.25)
Dunlee
(Picker / Philips)
PX1400 series
4” anode
“Q” stator
(6/12 Ω)
60 Hz
H.S.
SHIFT
CAP.
N/A
L.S.
SHIFT
CAP.
30µF
to
32µF
INDICO 100®
DSS PART NO.
(See note 2)
733317-13
735925-13
906894-13
LOW SPEED OPERATION
ONLY (See note 1)
HIGH SPEED OPERATION
ONLY (See note 1)
None
60 Hz
N/A
37.5µF
180 Hz
12.5µF
37.5µF
60/180 Hz
12.5µF
37.5µF
733317-16
735925-16
906894-16
733317-17
735925-17
906894-17
None
60/180 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 1
(10000)
None
60/180 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 13
(10110)
None
60/180 Hz
20 µF
60 µF
to
62 µF
733317-02
735925-02
906894-02
Switches
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
INDICO IQ®
DSS PART NO.
(See note 2)
904556-13
903255-13
904925-21
904925-22
904925-23
904925-21
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-13
901298-13
906672-21
906672-22
906672-23
901297-16
901298-16
906672-21
906672-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
904925-22
901297-02
901298-02
906672-22
Page 44
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
H.S.
SHIFT
CAP.
L.S.
SHIFT
CAP.
INDICO 100®
DSS PART NO.
(See note 2)
INDICO IQ®
DSS PART NO.
(See note 2)
CMP 200®
DSS PART NO.
(See note 2)
Type 14
(01110)
None
60/180 Hz
7.5
µF
36 µF
to
37.5µF
733317-12
735925-12
906894-12
904556-12
903255-12
904925-21
901297-12
901298-12
906672-21
Type 20
(00101)
None
60/180 Hz
7.5
µF
36 µF
to
37.5µF
733317-12
735925-12
906894-12
904556-12
903255-12
904925-21
901297-12
901298-12
906672-21
AR11-30
AR30-60
Type 22
(01101)
LOW SPEED
OPERATION ONLY
(See note 1)
50 Hz
N/A
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
AR20-50
AR30-100
AR40-100
Type 22
(01101)
None
50/150 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Eureka (see Varex/Varian)
GE Maxiray 75
(3” anode)
23/23Ω equal
impedance “E”
stator
* Dunlee tube
GE Maxiray 100
(4” anode)
23/23 Ω equal
impedance “E”
stator
Gilardoni
Rotagil
(33/36 Ω stator)
Gilardoni
Rotagil
A/A5
(33/36 Ω stator)
MX-75
DU303 (1.0/2.0)
*
DU303 (0.6/1.2)
*
MX-100
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 45
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
50 Hz
H.S.
SHIFT
CAP.
N/A
L.S.
SHIFT
CAP.
30 µF
to
36 µF
IAE
C30
C31
(25/62 Ω stator)
X20P
X22
X40
RTM70H
RTM 70HS
Type 10
(01010)
LOW SPEED
OPERATION ONLY
(See note 1)
Type 10
(01010)
Additional boost time may be
required at 400VAC (See note
12)
LOW SPEED
OPERATION ONLY
(See note 1)
50/150 Hz
6 µF
60 Hz
N/A
30 µF
to
36 µF
20 µF
to
30 µF
IAE
C40
(33/64 Ω stator)
X39
X40
X42
Type 19
(11001)
IAE
C52
C52 Super
C100
C352
(20/40 Ω stator)
RTM 78H
RTM 90H
RTM 92H
RTM 101H
RTM 102H
RTM 780H
RTM 782H
X39 X40
X42 X50H
X50AH
RTC 600HS
RTC 700HS
RTC 1000HS
RTM 78HS
RTM 90HS
RTM 92HS
RTM 101HS
RTM 102HS
RTM 780HS
RTM 782HS
Type 27
(11011)
LOW SPEED
OPERATION ONLY
(See note 1)
50 Hz
N/A
30 µF
to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Type 27
(11011)
Additional boost time may be
required at 400VAC
(See note 12)
50/150 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
733317-17
735925-17
906894-17
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
INDICO 100®
DSS PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
Rev. AF
Page 46
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT FREQ.
Kailong RADII
H2100
(20/50 Ω stator)
KL100
Type 21
(10101)
None
Machlett (Varex)
Dynamax 52
Std “R” stator
(16/50 Ω)
A-102
A-132
A-142
Type 0
(00000)
Machlett (Varex)
Dynamax 62
“STD” stator
(23/56 Ω)
Dynamax 62U
configured as
“STD” or “R”
(15/36 Ω)
A-192B
A-196
A-197
Type 0
(00000)
A-256
A-272
A-282
A-292
A-482
Type 1
(10000)
A-286
Switches
(1........5)
60/180Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF
to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
None
60/180 Hz
6 µF
30 µF
to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
None
60/180 Hz
6 µF
30 µF
to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
A-486
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 47
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
Philips
ROT 350
ROT 351
(36/44 Ω stator)
(See note 4)
RO 1230
RO 1750
RO 2050
RO 3050
SRO 2250
Type 9
(10010)
LOW SPEED
OPERATION ONLY
(See note 1)
Type 9
(10010)
* Dunlee tube
(See note 6)
** NAGO tube
*** Varex
(Varian)
tube
Philips
ROT 350
ROT 351
(9/11 Ω stator)
(See note 5)
* Dunlee tube
(See note 6)
*** Varex
(Varian)
tube
Philips
ROT 500
ROT 501
(9/11 Ω stator)
60 Hz
H.S.
SHIFT
CAP.
N/A
L.S.
SHIFT
CAP.
30µF to
32µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-13
901298-13
906672-21
906672-22
906672-23
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
37.5µF
904925-21
904925-23
901297-16
901298-16
906672-21
906672-23
Type 9
(10010)
None
60/180 Hz
12.5µF
37.5µF
733317-16
735925-16
906894-16
733317-17
735925-17
906894-17
PG-256 ***
PG-292 ***
RAD-34 ***
RAD-50 ***
RO 2550
RO 33100
SRO 1330
DU 2550 *
DU 33100 *
SRM 0310
Type 9
(10010)
HIGH SPEED
OPERATION ONLY
(See note 1)
180 Hz
12.5µF
37.5µF
Type 30
(01111)
None
60/180 Hz
12.5µF
37.5µF
904925-21
904925-23
SRM 1080
SRM 35100
Type 30
(01111)
HIGH SPEED
OPERATION ONLY
(See note 1)
180 Hz
733317-16
735925-16
906894-16
733317-17
735925-17
906894-17
901297-16
901298-16
906672-21
906672-23
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
DU 1750 *
DU 3050 *
GD6 3050 **
PG-256 ***
PG-292 ***
RAD-34 ***
RAD-50 ***
RO 1230
RO 1750
RO 3050
DU 1750 *
DU 3050 *
RO 2050
SRO 2250
37.5µF
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 48
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
H.S.
SHIFT
CAP.
L.S.
SHIFT
CAP.
INDICO 100® DSS
PART NO.
(See note 2)
INDICO IQ®
DSS PART NO.
(See note 2)
CMP 200®
DSS PART NO.
(See note 2)
Shimadzu
(Circlex)
RX80
RX81
RX82
(7.5/35 Ω stator)
P18DE-85
Type 16
(00001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF
733317-13
735925-13
906894-13
P33D
P38D
P38DE-85
Type 16
(00001)
HIGH SPEED
OPERATION ONLY
(See note 1)
180 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
Shimadzu
(Circlex)
RX100
P38E
Type 16
(00001)
None
60/180 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
Shimadzu
(Circlex)
RX150
(8.4/24 Ω stator)
P18DK
Type 16
(00001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF
733317-13
735925-13
906894-13
P38DK
P324DK
Type 16
(00001)
HIGH SPEED
OPERATION ONLY
(See note 1)
180 Hz
5 µF
30 µF
904556-13
903255-13
904925-21
904925-22
904556-13
903255-13
904925-21
904925-22
904556-13
903255-13
904925-21
904925-22
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
901297-13
901298-13
906672-21
906672-22
901297-13
901298-13
906672-21
906672-22
901297-13
901298-13
906672-21
906672-22
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Picker (see Dunlee)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 49
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
Siemens
Biangulix
“S” stator
(1-phase only)
(14/18 Ω)
Siemens Opti150
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian)
tube
Siemens
OptiTop
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian)
tube
Siemens Optilix
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian)
tube
Siemens Megalix
Siemens SV 125
“S” stator
(1-phase only)
(14/18 Ω)
BI 125/20/40
BI 125/30/50
BI 150/30/50
BI 150/30/52R
150/30/52R
150/40/72C
150/12/50
150/40/80
150/30/50
150/40/102C
SG-796B *
150/40/80HC-100L
150/40/80HC-102L
OPERATING NOTES
OUTPUT
FREQ.
Type 23
(11101)
HIGH SPEED
OPERATION ONLY
(See note 1)
Type 23
(11101)
HIGH SPEED
OPERATION ONLY
(See note 1)
Switches
(1........5)
Type 23
(11101)
L.S.
SHIFT
CAP.
30 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-13
903255-13
904925-21
904925-22
CMP 200®
DSS PART NO.
(See note 2)
901297-13
901298-13
906672-21
906672-22
150 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
HIGH SPEED
OPERATION ONLY
(See note 1)
150 Hz
5 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
30 µF
Type 23
(11101)
HIGH SPEED
OPERATION ONLY
(See note 1)
150 Hz
5 µF
HIGH SPEED
OPERATION ONLY
(See note 1)
30 µF
30 µF
None
Type 8
(00010)
30 µF
30 µF
None
SG-256 *
125/30/82CM-120LW
125/40/82CM-120LW
150 Hz
H.S.
SHIFT
CAP.
5 µF
None
SG-1096 *
150/30/50C-100L
150/40/80C-100L
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
150 Hz
6 µF
30 µF
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 50
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
Siemens SV 150
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian)
tube
SV 150/30/50C100L
Type 23
(11101)
HIGH SPEED
OPERATION ONLY
(See note 1)
150 Hz
Siemens SV 150
“S” stator
(1-phase only)
(14/18 Ω)
* Varex (Varian)
tube
SV 150/40/80C100L
Siemens (SXVT)
RAY-6_1
RAY-8(S)_1
RAY-12(S)_1
(1-phase only)
(20/50 Ω stator)
Switches
(1........5)
SG-296B *
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
H.S.
SHIFT
CAP.
5 µF
L.S.
SHIFT
CAP.
30 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-13
903255-13
904925-21
904925-22
CMP 200®
DSS PART NO.
(See note 2)
901297-13
901298-13
906672-21
906672-22
30 µF
None
Type 23
(11101)
None
60/180 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
SV 150/12/40-S
SV 150/25/48-S
SV 150/22/54-S
Type 31
(11111)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
904925-23
901297-13
901298-13
906672-21
906672-22
906672-23
Siemens (SXVT)
RAY-13(S)_1
RAY-14(S)_1
(1-phase only)
(20/50 Ω stator)
SV 150/33/78R-S
Type 31
(11111)
None
60/180 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
Siemens (SXVT)
SDR
(1-phase only)
(20/50 Ω stator)
SDR 150/30/50-1
Type 31
(11111)
None
60/180 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
Siemens (SXVT)
SV 150
“S” stator
(1-phase only)
(14/18 Ω)
SV 150/40/80C100_1
Type 23
(11101)
None
150 Hz
5 µF
30 µF
733317-13
735925-13
906894-13
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
SG-796B *
(See note 7)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 51
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Toshiba
Rotanode
Most Toshiba x-ray tubes are identified by their “E” number, which is used to represent both the Tube Insert and the complete assembly. The
housing number is usually not shown on the identification label. A specific “E” number may be available with several different stator types, each
having different starter requirements. Furthermore, a specific stator type may require different starter requirements depending on the housing
used. For ease of use, the Toshiba tubes are sorted by Tube Type (“E” number) within this table. The installer must confirm the stator type for the
planned tube(s). If the stator type is not identified within the Toshiba documentation, match the stator winding resistance to the appropriate table
entry. Contact the factory if the desired Tube Type and stator type / winding resistance are not listed on the same line within this table.
DRX-3624H
30 µF to
733317-12
Type 17
None
60/180 Hz 6 µF
904556-12
901297-12
DRX-3724H
735925-12
903255-12
901298-12
36 µF
(10001)
DRX-4634H
906894-12
901297-13
904556-13
901298-13
733317-13
903255-13
735925-13
906672-21
904925-21
904925-22
906894-13
906672-22
904925-23
906672-23
30 µF to
733317-12
E7100
Type 17
None
60/180 Hz 6 µF
901297-12
904556-12
(See note 8)
735925-12
901298-12
903255-12
36 µF
(10001)
906894-12
904556-13
901297-13
733317-13
903255-13
901298-13
735925-13
904925-21
906672-21
906894-13
904925-22
906672-22
904925-23
906672-23
20 µF to
LOW SPEED
E7132
Type 19
60 Hz
N/A
733317-13
904556-13
901297-13
OPERATION ONLY
E7239
735925-13
903255-13
901298-13
30 µF
(11001)
(See note 1)
E7240
906894-13
904925-21
906672-21
904925-22
E7242
906672-22
(See note 8)
904925-23 * 906672-23 *
* Preferred
* Preferred
Toshiba
XH-112V-2
XH-112W
XS-AA stator
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-112V
XS-AG stator
(9.4/28.3 Ω)
Toshiba
XH-121
XH-126
XH-150
XS-AV stator
XS-RA stator
(27.5/58 Ω)
Toshiba
XH-112V
XS-AG stator
(9.4/28.3 Ω)
E7250
(See note 8)
CODE,
Switches
(1........5)
Type 17
(10001)
OPERATING NOTES
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
None
OUTPUT
FREQ.
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Rev. AF
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Page 52
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
Toshiba
XH-106V
XH-180
XH-181
XS-AL stator
(9.4/28.3 Ω)
E7252
(See note 8)
Type 29
(10111)
None
Toshiba
XH-106V
XH-180
XH-181
XS-RA stator
(27.5/58 Ω)
E7252
(See note 8)
Type 19
(11001)
Switches
(1........5)
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF
to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Dual Speed Starters “-15”
must be jumper-configured
to provide the required 28
µF low speed shift
capacitance.
Please refer to note 10
before attempting to use
this Dual Speed Starter
configuration.
LOW SPEED
OPERATION ONLY
(See note 1)
60/180 Hz
3 µF
23 µF
to
28 µF
733317-15
735925-15
906894-15
(See note 10)
60 Hz
N/A
20 µF
to
30 µF
733317-13
735925-13
906894-13
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Toshiba
XH-157
XS-RB stator
(20.2/38 Ω)
E7254
E7255
(See note 8)
Type 21
(10101)
None
60/180 Hz
6 µF
30 µF
to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Toshiba
XH-121
XH-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
E7299
(See note 8)
Type 19
(11001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
20 µF
to
30 µF
733317-13
735925-13
906894-13
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-15
901298-15
(See note 10)
906672-23
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
Page 53
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Toshiba
XH-106V
XH-180
XS-AL stator
(9.4/28.3 Ω)
E7813
(See note 8)
Type 29
(10111)
None
Toshiba
XH-121
XS-BA stator
(18/47.5 Ω)
E7843
(See note 8)
Type 0
(00000)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Toshiba
XH-121 XH-126
XH-150
XS-RA stator
(27.5/58 Ω)
E7861
(See note 8)
Type 19
(11001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
20 µF to
30 µF
733317-13
735925-13
906894-13
Toshiba
XH-112V
XS-AG stator
(9.4/28.3 Ω)
E7864
(See note 8)
Type 17
(10001)
None
60/180 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Toshiba
XH-121
XH-126
XS-AV stator
(27.5/58 Ω)
E7865
(See note 8)
Type 19
(11001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
20 µF to
30 µF
733317-13
735925-13
906894-13
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
Page 54
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF
to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Toshiba
XH-112V
XS-AG stator
(9.4/28.3 Ω)
E7869
(See note 8)
Type 17
(10001)
None
Toshiba
XH-121
XS-AV stator
XS-RA stator
(27.5/58 Ω)
E7876
(See note 8)
Type 19
(11001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
20 µF
to
30 µF
733317-13
735925-13
906894-13
Toshiba
XH-121
XH-126
XH-150
XS-AL stator
(9.4/28.3 Ω)
E7884
(See note 8)
Type 29
(10111)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF
to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
Toshiba
XH-121
XS-126
XS-AV stator
XS-RA stator
(27.5/58 Ω)
E7886
(See note 8)
Type 19
(11001)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
20 µF
to
30 µF
733317-13
735925-13
906894-13
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
904556-13
903255-13
904925-21
904925-22
904925-23 *
* Preferred
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
901297-13
901298-13
906672-21
906672-22
906672-23 *
* Preferred
Page 55
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
Varex (Varian)
B-100
“R” stator
(16/50 Ω)
A-102
A-132 / A-134*
A-142 / A-144*
* (See note 9)
A-145
A-102
A-132 / A-134*
A-142 / A-144*
* (See note 9)
Type 11
(11010)
Varex (Varian)
B-100
“Q” stator
(8/15 Ω)
OPERATING NOTES
OUTPUT
FREQ.
Type 0
(00000)
None
Type 2
(01000)
LOW SPEED
OPERATION ONLY
(See note 1)
None
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
60 Hz
N/A
60/180 Hz
20 µF
60 µF
to
62 µF
733317-02
735925-02
906894-02
904925-22
901297-02
901298-02
906672-22
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 56
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Varex (Varian)
B-130
B-130H
B-135H
B-146
B-147
B-150
B-150H
“R” stator
(16/50 Ω)
A-152
A-182 / A-184*
A-192 / A-194*
A-195
A-196
A-197
* (See note 9)
A-272 / A-274*
A-277 / A-278*
A-282 / A-284 *
A-286
A-292 / A-294*
A-482
G-242 G-256
G-292 / G-294*
G-296
G-892 G-896
* (See note 9)
A-152
A-182 / A-184*
A-192 / A-194*
A-195 / A-196
A-197 / A-252
A-272 / A-274*
A-277 / A-278*
A-282 / A-284 *
A-286
A-292 / A-294*
A-482
G-242 G-256
G-292 / G-294*
G-296
RAD-60
* (See note 9)
Type 0
(00000)
None
None
60/180 Hz
20 µF
60 µF
to
62 µF
733317-02
735925-02
906894-02
904925-22
901297-02
901298-02
906672-22
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Varex (Varian)
B-130
B-130H
B-135H
B-150
B-150H
“Q” stator
(6/11 Ω)
Switches
(1….....5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Type 1
(10000)
Type 12
(00110)
Page 57
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
Varex (Varian)
B-160
B-160H
B-165H
“R” Stator
(16/50 Ω)
G-1077
G-1078
G-1080
G-1082 / G1084*
G-1086 / G1087* G-1092 /
G-1094*
* (See note 9)
G-1582
G-1592
G-1593
Type 15
(11110)
None
Type 15
(11110)
SG-1096
Varex (Varian)
B-240H
“R” Stator
(14/46 Ω)
Varex
(Varian / Eureka)
Diamond
Emerald
LEO
“R” stator
(20/50 Ω)
Varex (Varian)
B-180
B-180H
B-185
B-185H
“R” Stator
(16/50 Ω)
Varex (Varian)
B-199
“S” Stator
(15/18 Ω)
60/180 Hz
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
None
60/180 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Type 18
(01001)
None
50/150 Hz
5 µF
30 µF
904556-13
903255-13
904925-21
904925-22
901297-13
901298-13
906672-21
906672-22
G-2090
Type 6
(01100)
LIMITED USAGE –
CONTACT FACTORY
100 /
180 Hz
6 µF
12.5µF
to
20µF
733317-13
735925-13
906894-13
733317-17
735925-17
906894-17
733317-16
735925-16
906894-16
904925-21
904925-22
904925-23
RAD-8
RAD-12
RAD-74
Type 0
(00000)
LOW SPEED
OPERATION ONLY
(See note 1)
60 Hz
N/A
30 µF to
36 µF
RAD-13
RAD-14
RAD-68
Type 0
(00000)
None
60/180 Hz
6 µF
30 µF to
36 µF
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
901297-16
901298-16
906672-21
906672-22
906672-23
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 58
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE 2: TUBE TYPES - HIGH SPEED STARTER (Continued)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE,
OPERATING NOTES
OUTPUT
FREQ.
Varex
(Varian / Eureka)
Sapphire
“R” stator
(20/50 Ω)
RAD-21
RAD-40
RAD-44
RAD-56
RAD-60 *
RAD-92
RAD-94
Type 5
(10100)
None
60/180
Hz
* For “Q” stator,
refer to B-130
housing
Switches
(1........5)
(SEE TABLE 3 FOR OPERATING
PARAMETERS)
H.S.
SHIFT
CAP.
6 µF
L.S.
SHIFT
CAP.
30 µF to
36 µF
INDICO 100® DSS
PART NO.
(See note 2)
733317-12
735925-12
906894-12
733317-13
735925-13
906894-13
INDICO IQ®
DSS PART NO.
(See note 2)
904556-12
903255-12
904556-13
903255-13
904925-21
904925-22
904925-23
CMP 200®
DSS PART NO.
(See note 2)
901297-12
901298-12
901297-13
901298-13
906672-21
906672-22
906672-23
Note for Table 2
NOTE 1: Tube types designated as low speed only or high speed only must be
programmed for low speed only or high speed only operation. Refer to the section
Tube Selection within Chapter 3 of the applicable generator service manual for
details. For some tubes listed as high speed only, the starter may be capable of
low speed operation but the Manufacturer’s data sheet for the Insert lists high
speed only. In other cases, the limitation is due to operational limitations of the
Dual Speed Starter itself.
NOTE 2: Dual Speed Starter part numbers 733317-XX (Indico 100®), 904556-XX
(Indico IQ®) and 901297-XX (CMP 200® DR) are used in 400 VAC 3φ generators /
power supplies and in 480 VAC 3φ generators / power supplies with a line adjusting
transformer.
Dual Speed Starter part numbers 735925-XX (Indico 100®), 903255-XX (Indico IQ®)
and 901298-XX (CMP 200® DR) are used in 230 VAC 1φ generators / power supplies
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 59
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
and in 480 VAC 3φ direct input generators / power supplies (using no line adjusting
transformer).
Dual Speed Starter part numbers 906894-XX (Indico 100®), 904925-XX (Indico IQ®)
and 906672-XX (CMP 200® DR) are used in both 400 VAC and 480 VAC 3φ
generators. These RoHS-compliant Starters replace the existing 73317-XX /
735925-XX series (Indico 100®), 904556-XX / 903255-XX series (Indico IQ®) and
901297-XX / 901298-XX series (CMP 200® DR), and may provide a greater selection
of available shift capacitance.
Where more than one Dual Speed Starter part number is referenced for a
particular tube type, any of these (applicable to the generator and input voltage) is
acceptable. For 2-tube systems using different tube types, both tube types must be
considered. Where available, the XXXXXX-21 starter is generally preferred, unless
otherwise indicated in the tables.
The Dual Speed Starters provide a selection of shift capacitance values for both
high speed and low speed operation, as identified by the DSS part number suffix
(“tab type”). Unless otherwise specified, capacitor selection is made automatically
by relays located on the Dual Speed Starter board. The following table lists the
capacitance values provided, along with applicable notes.
Also, refer to TABLE: DSS BOARD AND SHIFT CAPACITANCE CONFIGURATIONS
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 60
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
TABLE: DSS BOARD AND SHIFT CAPACITANCE CONFIGURATIONS (NOTE 2 Continued)
DSS Part No.
(applicability)
HIGH-SPEED
SHIFT
CAPACITANCE
LOW-SPEED SHIFT
CAPACITANCE
COMMENTS
XXXXXX-01
(INDICO 100® only)
XXXXXX-02
(All except Indico IQ®)
XXXXXX-12
(All except Indico IQ®)
XXXXXX-13
(All except Indico IQ®)
XXXXXX-15
(All except Indico IQ®)
6 µF
31 µF
May be replaced by XXXXXX-21 in all configurations where available.
20 µF
40 µF
6 µF, 7.5 µF
36 µF, 37.5 µF
Typically used to support “Q” stator tubes.
May be replaced by XXXXXX-22 in all configurations, where available.
May be replaced by XXXXXX-21 in all configurations, where available.
5 µF, 6 µF
30 µF
May be replaced by XXXXXX-21 in all configurations, where available.
3 µF, 6 µF
Typically used to support CGR (GE) Statorix or Toshiba E7252 (XS-R/RA) tubes.
May be replaced by XXXXXX-23 in all configurations, where available.
Manual configuration required; see Note 10.
XXXXXX-16
(All except Indico IQ®)
6 µF, 12.5 µF
15.5 µF, 18.5 µF
or 28 µF, 31 µF
(manually selected)
12.5 µF, 31.5 µF,
37.5 µF
XXXXXX-17
(INDICO 100® only)
5 µF, 12.5 µF
30 µF, 37.5 µF
XXXXXX-21
(All except Indico 100®)
5 µF, 6 µF,
7.5 µF, 12.5 µF
5 µF, 6 µF,
20 µF
3 µF, 6 µF,
12.5 µF
12.5 µF, 30 µF,
37.5 µF
20 µF, 30 µF,
62.5 µF
23 µF, 32.5 µF
XXXXXX-22
(All except Indico 100®)
XXXXXX-23
(All except Indico 100®)
Used to support Philips ROT350/500 tubes. May be used in place of Dual Speed
Starters “-12” or “-13” in two-tube systems that include both a standard “R” stator
tube (6 µF high speed shift capacitance) and a Philips ROT 350/500 tube.
All such configurations may be replaced by XXXXXX-21, where available.
Used in place of Dual Speed Starter “-13” in two-tube systems that include both a
standard “R” stator tube (6 µF of high speed shift capacitance) and an IAE tube
(5 µF of high speed shift capacitance). All such configurations may be replaced
by XXXXXX-21, where available. Do not use these starters in place of
XXXXXX-13 with Siemens tubes due to high-speed shift capacitor voltage
limitations.
Generally preferred where available, unless support may be required for Toshiba
E7252 tubes (see XXXXXX-23), unless otherwise indicated within the tables.
Typically used to support “Q” stator tubes.
Preferred over XXXXXX-02, where available.
Typically used where support may be required for CGR (GE) Statorix or Toshiba
E7252 (XS-R/RA) tubes. Preferred over XXXXXX-15, where available.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 61
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
NOTE 3: Comet tube inserts with the prefix “DI” and “DX” are interchangeable.
NOTE 4: Philips / Dunlee housing with the windings connected in series (low-speed
/ high impedance configuration).
NOTE 5: Philips / Dunlee housing with the windings connected in parallel (highspeed / low impedance configuration).
NOTE 6: Dunlee tube inserts with the prefix “DU” and Philips tube inserts with the
prefix “RO” are interchangeable. Select the corresponding Philips insert type within
the Generator software if the Dunlee type is not available.
NOTE 7: Siemens (SXVT) housing with the “phase-shifted” windings connected in
parallel (low impedance, high speed configuration).
NOTE 8: Complete Toshiba x-ray tube assemblies include the suffix “X, “FX”, “GX”,
or “JX”, which are electrically interchangeable.
NOTE 9: These X-ray tubes incorporate a control grid. Grid control is currently not
supported by CPI generators. Connect the grid connection to Cathode Common
when using these tubes, and select the insert type within the Generator software
corresponding to the equivalent non-grid tube.
NOTE 10: Dual Speed Starters 733317-15 / 735925-15 / 906894-15 (Indico 100) and
901297-15 / 901298-15 (CMP 200® DR) have jumper selectable phase shift
capacitance that allows selection of 15.5 µF or 28 µF low speed shift capacitors for
operation with CGR (GE) Statorix or Toshiba E7252 (XS-R/RA stator) tubes,
respectively, as per table 2. By default, this starter is factory set to the 28 µF
position. Confirm proper configuration per the section “Configuring Dual Speed
Starter XXXXXX-15” in chapter 2 before proceeding. These starters also support
standard “R” stators that require 6 µF of high speed shift capacitance.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 62
Table 2: Dual-Speed Starter Tube Select Table for Indico IQ® / Indio 100® / CMP 200® Generators
CPI Canada Inc
(INDICO 100® only): If configured for 28 µF of low speed capacitance, these Dual
Speed Starters also support tubes listed in Table 2 as requiring Dual Speed Starters
733317-13 / 735925-13 / 906894-15 and requiring 6 µF of high speed capacitance.
For two-tube operation, this configuration will support:
1. These tubes and the Toshiba E7252/XS-RA (high and low speed)
simultaneously, or;
2. These tubes and a CGR Statorix tube (high-speed only) simultaneously
3. The Toshiba E7252/XS-RA and a CGR tube simultaneously if one or both
of these tubes is operated in high-speed mode only.
(CMP 200® DR only): If configured for 28 µF of low speed capacitance, these Dual
Speed Starters also support tubes listed in Table 2 as requiring Dual Speed Starters
901297-13 / 901298-13 and requiring 6 µF of high-speed capacitance.
NOTE 11: When operating under 400 VAC low-line conditions, tubes requiring a
High Speed start voltage of 400 V or greater (refer to Table 3) may not be provided
with full power high-speed boost. In this case, the boost time may need to be
extended and the anode speed must be checked.
NOTE 12: The Tube Type (Code) indicated for this tube requires greater than 400
VAC line input for full power high-speed boost. When used with 400 VAC line, the
boost time may need to be extended and the anode speed must be checked.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 63
Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes
CPI Canada Inc
(This page intentionally left blank)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 64
Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes
CPI Canada Inc
Dual-Speed Starter and Indico IQ Low-Speed Starter X-ray Tube
Codes
Tube operating parameters for each tube type are selected by setting switches to
the applicable Code, as indicated in Table 1C (Indico IQ LSS) and Table 2 (DSS). For
reference, operating parameters applicable to each Code are contained in Table 3
below.
Reference notes are provided at the end of Table 3.
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 65
Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes
CPI Canada Inc
TABLE 3: TUBE TYPE CODES (DUAL SPEED STARTER and INDICO IQ® LOW SPEED STARTER)
TUBE
TYPE
(CODE)
CODE
Switches
(1........5)
H.S.
START
VOLTS
H.S.
RUN
VOLTS
H.S.
BOOST
TIME
H.S.
FREQ.
H.S.
BRAKE
VOLTS
H.S.
BRAKE
TIME
L.S.
START
VOLTS
L.S.
RUN
VOLTS
L.S.
BOOST
TIME
L.S.
FREQ.
Hz
60 Hz
H.S.
SHIFT
CAP.
6 µF
0
00000
400 V
100 V
1.4 sec
180 Hz
100 V
3.0 sec
240 V
60 V
1.4 sec
1
10000
400 V
100 V
1.9 sec
180 Hz
100 V
3.0 sec
240 V
70 V
2
01000
400 V
100 V
1.0 sec
180 Hz
100 V
3.0 sec
240 V
60 V
3
11000
400 V
100 V
5.0
sec*
180 Hz
150 V
3.0 sec
240 V
4
00100
420 V
70V
2.5 sec
180 Hz
170 V
3.0 sec*
5
10100
400 V
100 V
2.3 sec
180 Hz
100 V
6
01100
440 V
140 V
6.0
sec*
180 Hz
7
11100
240 V
120 V
2.3 sec
8
00010
400 V
90 V
9
10010
280 V
10
01010
11
-12, -13, -15, -16, -21, -22, -23
1.9 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
1.0 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
70 V
5.0 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
* ( 2.5 sec Ready)
240 V
50 V
1.6 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
* (LS braking 0.8 sec)
3.0 sec
240 V
50 V
2.3 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
250 V
4.0 sec**
400 V
110 V
5.0 sec
100
Hz
6 µF
***
-16, -21, -22, -23
* ( 5.0 sec Ready)
** (LS braking 2.0s)
*** (12.5µF-20µF L.S. cap.)
180 Hz
100 V
3.0 sec
240 V
50 V
2.3 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
3.0 sec
150 Hz
100 V
3.0 sec
240 V
100 V
1.5 sec
50 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
60 V
1.5 sec
180 Hz
80 V
3.0 sec
180 V
50 V
1.5 sec
60 Hz
12.5
µF
-12 (LS only), -16, -17, -21,
-22 (LS only), -23
440 V
100 V
0.8 sec
150 Hz
120 V
1.8 sec
220 V
50 V
0.6 sec
50 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
11010
290 V
70 V
0.8 sec
180 Hz
60 V
3.0 sec
150 V
50 V
0.8 sec
60 Hz
20 µF
-02, -22
12
00110
290 V
60/80V
1.3 sec
180 Hz
60 V
3.0 sec
150 /
180 V
50 V
1.3 sec
60 Hz
20 µF
-02, -22
13
10110
340 V
60 V
1.0 sec
180 Hz
100 V
3.0 sec
240 V
70 V
1.0 sec
60 Hz
20 µF
-02, -22
14
01110
400 V
90 V
0.9 sec
180 Hz
80 V
2.0 sec
230 V
70 V
0.9 sec
60 Hz
7.5 µF
-12, -21
15
11110
400 V
100 V
5.0 sec
180 Hz
150 V
3.0 sec
240 V
70 V
5.0 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
16
00001
340 V
80 V
1.6 sec
180 Hz
80 V
3.0 sec
150 V
50 V
1.6 sec
60 Hz
5 µF
-13, -17, -21, -22, -23 (LS only)
17
10001
320 V
80 V
2.3 sec
180 Hz
100 V
3.0 sec
200 V
50 V
1.5 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
18
01001
400 V
100 V
2.8 sec
150 Hz
120 V
3.0 sec
180 V
60 V
1.5 sec
50 Hz
5 µF
-13, -17, -21, -22, -23 (LS only)
19
11001
320 V
130 V
2.0 sec
180 Hz
130 V
3.0 sec
240 V
60 V
1.4 sec
60 Hz
3 µF
-12, -13, -16, -21,-22 (all LS only),
-15, -23 (LS/HS)
20
00101
400 V
90 V
1.0 sec
180 Hz
80 V
3.0 sec
230 V
70 V
1.0 sec
60 Hz
7.5 µF
-12, -21
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
APPLICABLE
DSS TYPES
(See note 1)
Page 66
Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes
CPI Canada Inc
TABLE 3: TUBE TYPE CODES (DUAL SPEED STARTER and INDICO IQ® LOW SPEED STARTER)
TUBE
TYPE
(CODE)
CODE
Switches
(1........5)
H.S.
START
VOLTS
H.S.
RUN
VOLTS
H.S.
BOOST
TIME
H.S.
FREQ.
H.S.
BRAKE
VOLTS
H.S.
BRAKE
TIME
L.S.
START
VOLTS
L.S.
RUN
VOLTS
L.S.
BOOST
TIME
L.S.
FREQ.
Hz
APPLICABLE
DSS TYPES
(See note 1)
60 Hz
H.S.
SHIFT
CAP.
6 µF
21
10101
420 V
100 V
2.2 sec
180 Hz
150 V
3.0 sec
240 V
80 V
1.8 sec
22
01101
340 V
60 V
1.4 sec
150 Hz
120 V
4.0 sec
220 V
60 V
23
11101
360 V
90 V
1.6 sec
150 Hz
80 V
3.0 sec
150 V
50 V
1.5 sec
50 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
1.6 sec
50 Hz
5 µF
-13, -21, -22, -23 (LS only)
24
00011
420 V
80 V
1.8 sec
150 Hz
150 V
3.0 sec
240 V
80 V
1.8 sec
50 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
25
10011
420 V
80 V
2.2 sec
150 Hz
150 V
3.0 sec
240 V
80 V
1.8 sec
50 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
26
01011
400 V
80 V
1.0 sec
150 Hz
120 V
27
11011
440 V
100 V
1.1 sec
150 Hz
120 V
3.0 sec
240 V
80 V
1.0 sec
50 Hz
5 µF
Reserved
1.8 sec
220 V
80 V
1.1 sec
50 Hz
5 µF
-13, -17, -21, -22, -23 (LS only)
28
00111
450 V
150 V
3.0 sec
150 Hz
120 V
3.0 sec
240 V
90 V
3.0 sec
50 Hz
3 µF
-15, -21 (LS only), -23
29
10111
240 V
60 V
2.2 sec
180 Hz
80 V
1.6 sec
130 V
50 V
1.3 sec
60 Hz
6 µF
-12, -13, -15, -16, -21, -22, -23
30
01111
420 V
50 V
1.8 sec
180 Hz
150 V
3.0 sec*
240 V
50 V
1.8 sec
60 Hz
12.5
µF
-16, -17, -21, -23
* (LS braking: 1.2 sec)
31
11111
330V
80 V
1.4 sec
180 Hz
70V
3.0 sec
240 V
80 V
1.4 sec
60 Hz
5 µF
-13, -21, -22, -23 (LS only)
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
-12, -13, -15, -16, -21, -22, -23
Page 67
Table 3: Dual-Speed Starter and Indico IQ® Low-Speed Starter X-ray Tube Codes
CPI Canada Inc
Notes and Warnings for Table 3
Warning:
The starter uses modulation strategies to obtain the desired
outputs. Measured voltages may not agree with those listed in the
table. However, the currents flowing in the stator windings are
equivalent to those that would exist if the stator was excited with
these voltages.
NOTE 1: Tube Type codes should only be used with the applicable DSS types
specified in this table (refer to Table 1 for applicable DSS Types).
Refer to Table 1 for Indico IQ LSS Part Numbers applicable to each Tube Type
(Housing and Insert).
Refer to Table 2 for DSS Part Numbers applicable to each Tube Type (Housing and
Insert).
Use and disclosure is subject to the restrictions on the back of the cover page of this CPI document.
X-ray Tube Stator Compatibility Tables
P/N: 746026-00
Rev. AF
Page 68